When a greenhouse humidity sensor reports 100% RH for several consecutive hours while air temperature is falling, the usual cause is condensation on the probe surface, not a calibration fault. Check the temperature and ventilation record first. Recalibration or replacement rarely fixes a wet sensing element.
A saturated humidity reading is physically possible in a greenhouse at night. What makes it suspicious is persistence combined with a falling temperature. A greenhouse humidity sensor that stays at 100% RH through a temperature drop of two or three degrees is describing its own surface, not the bulk air.
Relative humidity is referenced to saturation vapour pressure at the temperature of the sensing element. If that element is cooler than the surrounding air, the saturation pressure at the element is lower. The same quantity of water vapour then represents a higher percentage. Once the element reaches the dew point, liquid water forms.
The water film changes the local environment completely. The probe sits in contact with liquid instead of vapour, and the reading locks near 100% RH. It can remain there after the air has dried slightly, because evaporation from the film keeps the surface micro-layer close to saturation.
The practical rule is diagnostic order. Inspect the temperature trace and the ventilation schedule before you inspect a calibration certificate. A falling temperature with a pinned humidity channel points to condensation. A channel that drifts upward over weeks at stable temperature points somewhere else.
A single 100% RH sample is not evidence of anything. Night humidity rises naturally and short peaks are normal. The criterion requires duration: several hours of flat, ceiling-level readings while temperature declines is the pattern that justifies a condensation check.
True saturation means the bulk air holds all the vapour it can at its own temperature. Probe condensation means the air could hold more, but the sensing surface is cold enough to condense water out of it. The display looks identical. The physics, the recovery and the remedy do not.
Figure 1. Wall-mounted temperature and humidity sensors used in storage warehouses, greenhouses.
Temperature coupling. Compare the humidity trace with air temperature at the same logging interval. Condensation tracks the cooling curve; true saturation follows the air mass and can persist while temperature is flat.
Airflow response. Direct gentle airflow across the probe at about 0.6 m/s. A wet element falls several percentage points within minutes. Bulk air at saturation changes little, because moving saturated air does not dry it.
Side-by-side comparison. Place a second greenhouse humidity sensor at the same height, roughly 0.5 m away. Condensation often appears on one element only. Two probes at 100% RH in the same air suggests genuinely saturated air.
Vapour-pressure arithmetic. Convert the assumed bulk state to dew point and compare with the surface temperature. If the surface needs only a small temperature depression to condense, the pinned reading is consistent with condensation.
Values in the table are worked assumptions or a typical engineering reference value, not NiuBoL specifications. Confirm any quoted specification against the current datasheet.
Illustrative calculation based on assumed site values, not a measured field record. It is included to make the decision testable, not to describe a particular installation.
Input. Night air temperature 18 °C. The display reads 100% RH. Because the signature above suggests a wet element, assume the bulk air is actually at 90% RH rather than accepting the display. That 90% figure is an assumption.
Step 1 — dew point. Using the Magnus–Tetens relation, e_s(T) = 6.112 × exp(17.67 T / (T + 243.5)) hPa. At 18 °C, e_s = 20.6 hPa. Actual vapour pressure at 90% RH is 0.90 × 20.6 = 18.6 hPa. Solving for temperature gives a dew point of about 16.3 °C.
Step 2 — surface depression. The air is 18 °C and the dew point is 16.3 °C. The sensing surface therefore needs to be only 18 − 16.3 = 1.7 °C cooler than the air before water condenses on it. A night-time radiative loss of roughly 2 °C is enough.
Step 3 — recovery. Assume ventilation at 0.6 m/s and a first-order recovery time constant of 120 s, a typical engineering reference value rather than a published NiuBoL figure. The model is RH_read(t) = RH_true + (RH_start − RH_true) × exp(−t/τ).
With RH_true = 90%, RH_start = 100% and τ = 120 s, the time to come within one percentage point is τ × ln(10/1) = 120 × 2.303 = 276 s, about 4.6 min. The order of magnitude is minutes, not seconds and not hours.
Boundary. In still air the time constant lengthens considerably, so recovery can stretch well beyond ten minutes. Inside a dense canopy the air at leaf height may genuinely sit near 100% RH, in which case ventilation will not pull the reading down. Neither case is a calibration failure.
Co-location test. Keep the suspect greenhouse humidity sensor in place. Put a second unit at the same height and roughly 0.5 m away, with both logged at a 5 min interval. If one channel falls after ventilation while the other stays pinned, the fault is local to the first element. If both stay pinned, examine the air itself.
Figure 2. Leaf Temperature and Humidity Sensor For Smart Agriculture.
Ventilation test. Move air across the probe at about 0.6 m/s for five minutes and log at a short interval. A reading that drops several percentage points and then stabilises confirms a surface film. A reading that barely moves supports true saturation. Use a handheld anemometer to record the actual speed rather than assuming it.
Both checks are reversible and take no laboratory. They also keep the installation intact, which matters because moving a probe changes its thermal history and can create a new condensation event. Record the start time of ventilation so the recovery curve can be compared with the model above.
For logging, RS485 Modbus RTU preserves the raw humidity word alongside temperature. That is useful here because the diagnostic depends on the relationship between two channels. A 4–20 mA loop can hide channel identity inside the current, so confirm the register map and scaling on the current datasheet.
A greenhouse humidity sensor measures the air immediately around its sensing element. Inside a crop canopy that air is not the same as the air in the walkway or above the crop. Transpiration adds vapour, and leaf layers slow mixing. Near-saturation at leaf height can be real.
Height is therefore part of the diagnosis. A probe mounted among dense leaves at night may sit in a microclimate that differs from a reference several metres away. Before calling the reading an error, compare with a second point at a different height and record both temperatures.
Avoid contact with wet foliage. A leaf touching the housing transfers liquid water directly and produces the same pinned 100% RH signature as internal condensation. Keep clearance of at least a few centimetres and note the distance in the commissioning record.
Sunlight, irrigation mist and door openings all change the local state quickly. Note them in the log, because a recovery curve that ignores a misting cycle cannot be interpreted. None of these notes require a new instrument; they require a consistent logging convention.
The method assumes the reading is caused by water on the element. It fails when the air is genuinely saturated for hours, as in fog, after rain or inside a closed, dense canopy. In that state both co-located probes remain at 100% RH and ventilation changes little. The correct response is to accept the reading and improve ventilation of the space.
It also fails if the element has been water-damaged. A sensor that has stayed wet for long periods may recover slowly or not at all, and a longer time constant does not prove a healthy instrument. Check the manufacturer's stated operating humidity range and replacement guidance.
The temperature-difference test is weak when the probe sits in still, shaded air with no radiative cooling. In that case a small surface depression is not expected, and a pinned reading needs a different explanation. Rising temperature with 100% RH also falls outside this pattern.
Finally, this diagnostic says nothing about root-zone water or crop condition. It is a check on an air humidity measurement and its mounting. Do not extend it into irrigation scheduling or production forecasts.
Q1. Why does my greenhouse humidity sensor stick at 100% RH overnight?
Usually because the sensing surface is cooler than the air and has collected liquid water. Several hours at 100% RH while temperature falls by 2–4 °C is the classic signature. Check the temperature trace before considering recalibration.
Figure 3. Measures air temperature and relative humidity pressure Sensor.
Q2. How can I tell condensation from genuinely saturated air?
Blow air across the probe at about 0.6 m/s for five minutes. A wet element drops several percentage points within minutes; truly saturated air barely changes. A second probe 0.5 m away provides an independent comparison.
Q3. How long does recovery take after ventilation starts?
With an assumed time constant of 120 s at 0.6 m/s, a reading moving from 100% to a true 90% RH comes within one percentage point after about 276 s, or 4.6 min. That constant is a typical engineering reference value.
Q4. At 18 °C, how much colder must the probe be to condense?
If the bulk air is assumed at 90% RH, dew point is about 16.3 °C. The surface needs to be roughly 1.7 °C below the 18 °C air. A 2 °C night-time radiative loss is enough.
Q5. Should I replace the sensor when it reads 100% RH?
Not on the reading alone. Run the co-location and ventilation checks first. Replacement is justified only if the unit fails both checks, stays wet after long ventilation, or falls outside the stated operating range.
Q6. Can a high night reading be correct?
Yes. Inside a dense canopy, transpiration and poor mixing can hold air near 100% RH for hours. Both probes then read the same value and ventilation changes little. Treat this as a real microclimate, not a fault.
A greenhouse humidity sensor pinned at 100% RH during a temperature fall is most often reporting condensation on its own surface. The decisive evidence is temperature coupling, a drop under ventilation, and disagreement with a co-located probe. The worked example puts dew point near 16.3 °C at 18 °C air and recovery on a minutes scale.
Before trusting the next night's data:
Read temperature before you touch calibration
A greenhouse humidity sensor parked at 100% RH may be measuring condensation on its own surface. At 18 °C and 90% RH the dew point is about 16.3 °C, so roughly 1.7 °C of radiative cooling is enough to form water on the probe; with a 120 s recovery constant, returning within one percentage point takes about 276 s. Compare with a ventilated reference and check the temperature trace before you replace or recalibrate the probe.
Seeing a humidity trace pinned at 100% RH? Send the air temperature, probe location and ventilation notes on our contact page, and we will help you separate condensation from bias.
Prev:Turbidity Sensor: Why Bubbles Change the NTU Reading
Next:no more
Related recommendations
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related products
Combined air temperature and relative humidity sensor
TDR Soil Moisture & Temperature Sensor for irrigation| RS485 |4-20mA| NBL-S-THR
Soil pH sensor RS485 soil Testing instrument soil ph meter for agriculture |NBL-S-PH
Wind Speed sensor Output Modbus/RS485/Analog/0-5V/4-20mA
0.2mm Tipping Bucket Rain Gauge for Weather & Hydrology | NiuBoL
Pyranometer Solar Radiation Sensor 4-20mA/RS485
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)