Rain Gauge Check: Why 20 mm of Water Reads 18.6 mm

Time:2026-11-26 12:00:00 Popularity:5

A rain gauge is only as credible as its measurement chain: collecting area, bucket volume and logger scaling. With a 200 mm aperture and a 6 mL bucket, 628 mL of water should read about 20.0 mm. If the display shows 18.6 mm, the shortfall is 7.0% and the cause sits in that chain, not in the brand name.

The 628 mL hand check: what should a 200 mm aperture record?

Take a 200 mm collecting aperture and a 6 mL tipping bucket. The radius is 100 mm, so the area is π × 100² = 31,416 mm², or 314.16 cm². One tip empties 6 mL, which is 6,000 mm³. Dividing volume by area gives 6,000 / 31,416 = 0.191 mm per tip. That number is the geometric resolution of the rain gauge; no extra display digit can make it smaller.

Now inject 628 mL without splashing. The depth represented is 628,000 mm³ / 31,416 mm² = 19.99 mm, which rounds to 20.0 mm. The same result comes from tips: 628 / 6 = 104.67 tips, and 104.67 × 0.191 = 20.0 mm. Because a real tip is an integer, a clean run lands on 104 or 105 tips, or roughly 19.9–20.1 mm. The arithmetic does not depend on the sensor brand.

If the logger reports 18.6 mm, the relative deviation is (18.6 − 20.0) / 20.0 = −7.0%. Against the stated +/-4% tolerance at 2 mm/min, that sits outside the band. This is an illustrative calculation based on assumed site values, not a measured field record; it exists so you can reproduce the check on a bench with a known volume and a stopwatch.

Step Quantity Input / formula Result Unit / condition
1 Collecting diameter given 200 mm
2 Collecting radius diameter / 2 100 mm
3 Collecting area π × 100² 31,416 mm² (314.16 cm²)
4 Bucket volume given 6 mL (= 6,000 mm³)
5 Per-tip equivalent 6,000 / 31,416 0.191 mm per tip
6 Injected volume given 628 mL (= 628,000 mm³)
7 Theoretical tips 628 / 6 104.67 tips
8 Theoretical depth 628,000 / 31,416 19.99 mm (about 20.0 mm)
9 Logged depth assumed observation 18.6 mm
10 Absolute error 18.6 − 20.0 −1.4 mm
11 Relative deviation −1.4 / 20.0 −7.0 %
12 Test rain rate stated condition 2 mm/min
13 Tips at that rate 2 / 0.191 10.5 tips/min
14 Time per tip 60 / 10.5 5.7 s

The first, fourth, sixth, ninth and twelfth rows are assumed or stated inputs; every other row is derived from them. No field measurement is implied.

Aperture area and bucket volume are the only two constants

The millimetre depth reported by a rain gauge is not a sensor output in the usual sense. It is a volume divided by a horizontal collecting area. Change either constant and every reading scales with it. A 6 mL bucket over a 314.16 cm² aperture yields 0.191 mm per tip; a 6.28 mL bucket over the same aperture yields 0.200 mm per tip. The NiuBoL guide specifies 0.2 mm increments from a 200 mm aperture, which corresponds to about 6.28 mL per tip.

7-in-1 Ultrasonic Weather Sensor with Rainfall & illuminance | NBL-W-71MUWS

Figure 1. 7-in-1 Ultrasonic Weather Sensor with Rainfall & illuminance — NBL-W-71MUWS.

That 0.009 mm difference is small, but it is systematic. If the logger is configured for 0.2 mm while the physical bucket delivers 0.191 mm, each tip is credited about 4.7% too much depth across a long event. Read the actual bucket volume and the logger coefficient as a pair, not as two independent settings.

Extra decimals do not add resolution. The NiuBoL guide notes that a logger showing 2.000 mm has gained no physical precision: ten completed tips represent 2 mm on the 0.2 mm configuration. Store the raw tip count next to the scaled depth so that any later scaling mistake can be reconstructed.

A controlled-water injection procedure you can repeat

  1. Confirm the physical constants. Read the aperture diameter and the full-tip bucket volume from the current datasheet, and write both into the test sheet before adding any water.
  2. Clear the collector path. Remove leaves, insects and webbing from the funnel and the bucket pivot; debris changes effective volume and can hold the bucket partway through its travel.
  3. Prepare known volumes. Use graduated vessels at 62.8 mL, 314 mL and 628 mL, all near room temperature. At 314.16 cm² these represent about 2.0 mm, 10.0 mm and 20.0 mm.
  4. Inject slowly and evenly. Feed each volume into the funnel over several minutes so the bucket tips one at a time. A fast pour creates splash loss and a false under-read.
  5. Count tips independently. Record the logger tip total and, if fitted, a separate counter. Compare each with the injected volume divided by bucket volume.
  6. Repeat and log conditions. Run at least three repeats per level, noting water and air temperature. Evaporation from the funnel during a slow injection is small but real.
  7. Calculate deviation. For each level compute (logged − theoretical) / theoretical × 100. A result outside the stated tolerance at the stated rain rate calls for a deeper check.
Injected volume (mL) Aperture area (cm²) Bucket volume (mL) Theoretical tips Whole tips logged Theoretical depth (mm)
6.28 314.16 6.00 1.05 1 0.20
31.4 314.16 6.00 5.23 5 1.00
62.8 314.16 6.00 10.47 10 2.00
125.6 314.16 6.00 20.93 20 4.00
314 314.16 6.00 52.33 52 10.00
628 314.16 6.00 104.67 104 20.00
628 314.16 6.28 100.00 100 20.00

The last row shows why the bucket volume matters: the same 628 mL gives 100 clean tips at 6.28 mL, but 104.67 non-integer tips at 6.00 mL. Only whole tips can be recorded.

Rain-rate ceiling: 2 mm/min and the mechanical response

A rain gauge also has a dynamic limit. The NiuBoL guide quotes +/-4% at an indoor test condition of 2 mm/min. At the 0.191 mm geometric equivalent, 2 mm/min means 2 / 0.191 = 10.5 tips per minute, or one tip every 5.7 seconds. On the nominal 0.2 mm configuration it is exactly 10 tips per minute, one every 6 seconds.

Whether 10.5 tips/min approaches the mechanism's response ceiling is a datasheet question, not a guess. A settle time well below one second is a typical engineering reference value for a small tipping bucket, so the 5.7 s interval leaves margin. Confirm the actual recovery time on the current datasheet before running a rain gauge above 2 mm/min.

Above the ceiling, two losses appear together. The funnel delivers water faster than the bucket can empty, so a new tip begins before the previous travel has ended. Water striking a moving or nearly full bucket also splashes out of the collector. Both are negative errors, so a rain gauge at high intensity tends to under-read rather than over-read.

Splash is not only a high-rate problem. A jet aimed directly at the bucket, or a funnel outlet that drops water onto the pivot instead of into the cup, scatters droplets outside the collecting path. During verification, pour along the funnel wall or use a diffuser.

From tipping pulse to logged millimetres

The bucket produces a contact closure, not a depth. The recorder multiplies each pulse by a coefficient. For the NBL-W-RS, ten completed tips represent 2 mm on the 0.2 mm configuration, and the guide gives three tips in five minutes as 0.6 mm with an interval-average rate of 7.2 mm/h. Check the arithmetic first: 3 × 0.2 = 0.6 mm, and 0.6 × 12 = 7.2 mm/h.

7-in-1 Compact Weather Sensor with Radar Rain Gauge | Ultrasonic Weather Station Sensor| RS485/SDI-12

Figure 2. 7-in-1 Compact Weather Sensor with Radar Rain Gauge — Ultrasonic Weather Station Sensor — RS485/SDI-12.

Do not multiply an already scaled rainfall value by 0.2 again. If the logger stores millimetres, applying the pulse coefficient a second time shrinks the record by a factor of five. If it stores raw pulses, apply the coefficient once at export. Keep the raw tip count in the archive so the conversion can be audited.

Voltage and digital outputs add a second scaling step. The guide lists 0–2.5 V and 0–5 V versions that use the matching voltage input and delivered conversion, alongside pulse and RS485 versions. A 0–5 V channel read on a 0–2.5 V input range doubles the interpreted depth. Cumulative error differs from single-tip error: a 1% coefficient mistake becomes 1% of every event, while a dropped pulse leaves a permanent 0.191–0.2 mm gap in the total.

For that reason, acceptance compares raw tips first and scaled depth second. A mismatch that appears only after scaling is a logger configuration fault; a mismatch in raw tips is a hydraulic or mechanical fault inside the rain gauge.

Where this check stops being valid

The controlled-water check validates the measurement chain under a narrow set of conditions. It does not validate rainfall capture. Evaporation from the funnel removes water before a tip occurs, and the effect grows with temperature, wind and long injection times. Splash-out removes water at impact. Neither appears in the tip count, and both bias the result low.

Bucket residue is the opposite case. A film of water clinging to the cup after a tip changes the volume needed for the next one, so successive tips are not identical. Let the bucket drain and inspect the pivot before testing. Dust, insect debris and spider webbing add volume and friction; clean them with water and a soft brush, then repeat the check.

The check also stops being valid outside the stated test point. The +/-4% figure is tied to 2 mm/min indoor conditions and is not a blanket field tolerance. Frozen or mixed precipitation, drizzle near the resolution threshold, and rates above the response ceiling all sit outside it. Use the result as a reproducibility test of the chain, not as a site acceptance certificate.

FAQ

Q1. How many millimetres should 628 mL produce on a 200 mm aperture?

About 20.0 mm. The area is π × 100² = 31,416 mm², and 628,000 mm³ / 31,416 mm² = 19.99 mm. With a 6 mL bucket that is 104.67 tips; with a 6.28 mL bucket it is exactly 100 tips. This is an illustrative calculation based on assumed site values, not a measured field record.

Q2. Is 0.191 mm or 0.2 mm the correct per-tip value?

Both can be right under different assumptions. A 6 mL bucket over 314.16 cm² gives 0.191 mm geometrically; the NiuBoL guide specifies 0.2 mm increments, which implies about 6.28 mL per tip. Read the physical bucket volume and the logger coefficient together, then apply the pair consistently.

Q3. What does a −7.0% deviation at 628 mL actually mean?

It means the logged 18.6 mm is 1.4 mm below the 20.0 mm theoretical depth. That is outside the stated +/-4% tolerance at 2 mm/min, so check splash loss, bucket residue, the injected volume and the pulse coefficient before drawing any conclusion about the rain gauge.

Q4. Does a logger showing 2.000 mm mean 0.001 mm resolution?

No. The NiuBoL guide notes that extra decimals do not improve physical resolution. Ten tips represent 2 mm on the 0.2 mm configuration, so the meaningful increment is 0.2 mm. Keep the raw tip count, which shows exactly which increments were recorded.

Q5. How many tips per minute is 2 mm/min?

At 0.191 mm per tip it is 2 / 0.191 = 10.5 tips/min, one every 5.7 seconds. At the nominal 0.2 mm it is 10 tips/min, one every 6 seconds. A settle time well below one second is a typical engineering reference value; verify it on the current datasheet.

Q6. Can I use this check for every rainfall type?

No. The injection check covers liquid water delivered slowly near room temperature. Evaporation, splash-out, bucket residue, frozen precipitation and rates above the response ceiling sit outside it. Treat the result as a chain verification, then repeat at several rates if you need a wider operating envelope.

Summary and next actions

A rain gauge converts water volume into depth through two constants: aperture area and bucket volume. A 200 mm aperture gives 314.16 cm²; a 6 mL bucket makes 0.191 mm per tip; 628 mL should read 20.0 mm. An 18.6 mm reading is a −7.0% deviation to be traced through splash, residue and logger scaling.

Portable Tripods Weather Station Stands

Figure 3. Portable Tripods Weather Station Stands.

  1. Write the aperture diameter, bucket volume and logger coefficient on the commissioning sheet.
  2. Inject 62.8, 314 and 628 mL in triplicate; log raw tips and scaled depth separately.
  3. Compute deviation at each level, then repeat after cleaning the funnel and bucket.
  4. Run a 2 mm/min equivalent pour and confirm the tip interval exceeds the datasheet recovery time.
  5. Archive raw tip counts beside each scaled record so any cumulative error can be reconstructed.

Verify the coefficient against the physical bucket

A rain gauge reports volume divided by aperture area, not a raw sensor output. A 200 mm aperture and a 6 mL bucket give 0.191 mm per tip, while the 0.2 mm setting implies about 6.28 mL; that small constant difference becomes roughly a 4.7% depth error over a long event. Inject a known 628 mL, compare the 20.0 mm expectation with the logger, and keep raw tip counts next to the scaled depth.

Validating a tipping bucket with a known water volume? Tell us the aperture, bucket size and logger coefficient on our contact page, and we will help you check the per-tip value.

Related recommendations

Sensors & Weather Stations Catalog

Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf

Weather Stations Catalog-NiuBoL.pdf

Agriculture Sensors Catalog-NiuBoL.pdf

Water Quality Sensor Catalog-NiuBoL.pdf

Related products

Tell us your requirements, Let's discuss more about your project.we can do more.

Name*

Tel*

Email*

Company*

Country*

Message


online
Contacts
Email
Top
XRain Gauge Check: Why 20 mm of Water Reads 18.6 mm-Product knowledge-Automatic Weather Stations_Industrial, Agricultural, Water & Environmental IoT Monitoring Solutions—NiuBoL

Screenshot, WhatsApp to identify the QR code

WhatsApp number:+8615367865107

(Click on WhatsApp to copy and add friends)

Open WhatsApp

The WhatsApp ID has been copied, please open WhatsApp to add consultation details!
WhatsApp