— Blogs —
—Products—
Consumer hotline +8618073152920 WhatsApp:+8615367865107
Address:Room 102, District D, Houhu Industrial Park, Yuelu District, Changsha City, Hunan Province, China
Product knowledge
Time:2026-11-26 12:00:00 Popularity:5
A rain gauge is only as credible as its measurement chain: collecting area, bucket volume and logger scaling. With a 200 mm aperture and a 6 mL bucket, 628 mL of water should read about 20.0 mm. If the display shows 18.6 mm, the shortfall is 7.0% and the cause sits in that chain, not in the brand name.
Take a 200 mm collecting aperture and a 6 mL tipping bucket. The radius is 100 mm, so the area is π × 100² = 31,416 mm², or 314.16 cm². One tip empties 6 mL, which is 6,000 mm³. Dividing volume by area gives 6,000 / 31,416 = 0.191 mm per tip. That number is the geometric resolution of the rain gauge; no extra display digit can make it smaller.
Now inject 628 mL without splashing. The depth represented is 628,000 mm³ / 31,416 mm² = 19.99 mm, which rounds to 20.0 mm. The same result comes from tips: 628 / 6 = 104.67 tips, and 104.67 × 0.191 = 20.0 mm. Because a real tip is an integer, a clean run lands on 104 or 105 tips, or roughly 19.9–20.1 mm. The arithmetic does not depend on the sensor brand.
If the logger reports 18.6 mm, the relative deviation is (18.6 − 20.0) / 20.0 = −7.0%. Against the stated +/-4% tolerance at 2 mm/min, that sits outside the band. This is an illustrative calculation based on assumed site values, not a measured field record; it exists so you can reproduce the check on a bench with a known volume and a stopwatch.
The first, fourth, sixth, ninth and twelfth rows are assumed or stated inputs; every other row is derived from them. No field measurement is implied.
The millimetre depth reported by a rain gauge is not a sensor output in the usual sense. It is a volume divided by a horizontal collecting area. Change either constant and every reading scales with it. A 6 mL bucket over a 314.16 cm² aperture yields 0.191 mm per tip; a 6.28 mL bucket over the same aperture yields 0.200 mm per tip. The NiuBoL guide specifies 0.2 mm increments from a 200 mm aperture, which corresponds to about 6.28 mL per tip.
Figure 1. 7-in-1 Ultrasonic Weather Sensor with Rainfall & illuminance — NBL-W-71MUWS.
That 0.009 mm difference is small, but it is systematic. If the logger is configured for 0.2 mm while the physical bucket delivers 0.191 mm, each tip is credited about 4.7% too much depth across a long event. Read the actual bucket volume and the logger coefficient as a pair, not as two independent settings.
Extra decimals do not add resolution. The NiuBoL guide notes that a logger showing 2.000 mm has gained no physical precision: ten completed tips represent 2 mm on the 0.2 mm configuration. Store the raw tip count next to the scaled depth so that any later scaling mistake can be reconstructed.
The last row shows why the bucket volume matters: the same 628 mL gives 100 clean tips at 6.28 mL, but 104.67 non-integer tips at 6.00 mL. Only whole tips can be recorded.
A rain gauge also has a dynamic limit. The NiuBoL guide quotes +/-4% at an indoor test condition of 2 mm/min. At the 0.191 mm geometric equivalent, 2 mm/min means 2 / 0.191 = 10.5 tips per minute, or one tip every 5.7 seconds. On the nominal 0.2 mm configuration it is exactly 10 tips per minute, one every 6 seconds.
Whether 10.5 tips/min approaches the mechanism's response ceiling is a datasheet question, not a guess. A settle time well below one second is a typical engineering reference value for a small tipping bucket, so the 5.7 s interval leaves margin. Confirm the actual recovery time on the current datasheet before running a rain gauge above 2 mm/min.
Above the ceiling, two losses appear together. The funnel delivers water faster than the bucket can empty, so a new tip begins before the previous travel has ended. Water striking a moving or nearly full bucket also splashes out of the collector. Both are negative errors, so a rain gauge at high intensity tends to under-read rather than over-read.
Splash is not only a high-rate problem. A jet aimed directly at the bucket, or a funnel outlet that drops water onto the pivot instead of into the cup, scatters droplets outside the collecting path. During verification, pour along the funnel wall or use a diffuser.
The bucket produces a contact closure, not a depth. The recorder multiplies each pulse by a coefficient. For the NBL-W-RS, ten completed tips represent 2 mm on the 0.2 mm configuration, and the guide gives three tips in five minutes as 0.6 mm with an interval-average rate of 7.2 mm/h. Check the arithmetic first: 3 × 0.2 = 0.6 mm, and 0.6 × 12 = 7.2 mm/h.
Figure 2. 7-in-1 Compact Weather Sensor with Radar Rain Gauge — Ultrasonic Weather Station Sensor — RS485/SDI-12.
Do not multiply an already scaled rainfall value by 0.2 again. If the logger stores millimetres, applying the pulse coefficient a second time shrinks the record by a factor of five. If it stores raw pulses, apply the coefficient once at export. Keep the raw tip count in the archive so the conversion can be audited.
Voltage and digital outputs add a second scaling step. The guide lists 0–2.5 V and 0–5 V versions that use the matching voltage input and delivered conversion, alongside pulse and RS485 versions. A 0–5 V channel read on a 0–2.5 V input range doubles the interpreted depth. Cumulative error differs from single-tip error: a 1% coefficient mistake becomes 1% of every event, while a dropped pulse leaves a permanent 0.191–0.2 mm gap in the total.
For that reason, acceptance compares raw tips first and scaled depth second. A mismatch that appears only after scaling is a logger configuration fault; a mismatch in raw tips is a hydraulic or mechanical fault inside the rain gauge.
The controlled-water check validates the measurement chain under a narrow set of conditions. It does not validate rainfall capture. Evaporation from the funnel removes water before a tip occurs, and the effect grows with temperature, wind and long injection times. Splash-out removes water at impact. Neither appears in the tip count, and both bias the result low.
Bucket residue is the opposite case. A film of water clinging to the cup after a tip changes the volume needed for the next one, so successive tips are not identical. Let the bucket drain and inspect the pivot before testing. Dust, insect debris and spider webbing add volume and friction; clean them with water and a soft brush, then repeat the check.
The check also stops being valid outside the stated test point. The +/-4% figure is tied to 2 mm/min indoor conditions and is not a blanket field tolerance. Frozen or mixed precipitation, drizzle near the resolution threshold, and rates above the response ceiling all sit outside it. Use the result as a reproducibility test of the chain, not as a site acceptance certificate.
Q1. How many millimetres should 628 mL produce on a 200 mm aperture?
About 20.0 mm. The area is π × 100² = 31,416 mm², and 628,000 mm³ / 31,416 mm² = 19.99 mm. With a 6 mL bucket that is 104.67 tips; with a 6.28 mL bucket it is exactly 100 tips. This is an illustrative calculation based on assumed site values, not a measured field record.
Q2. Is 0.191 mm or 0.2 mm the correct per-tip value?
Both can be right under different assumptions. A 6 mL bucket over 314.16 cm² gives 0.191 mm geometrically; the NiuBoL guide specifies 0.2 mm increments, which implies about 6.28 mL per tip. Read the physical bucket volume and the logger coefficient together, then apply the pair consistently.
Q3. What does a −7.0% deviation at 628 mL actually mean?
It means the logged 18.6 mm is 1.4 mm below the 20.0 mm theoretical depth. That is outside the stated +/-4% tolerance at 2 mm/min, so check splash loss, bucket residue, the injected volume and the pulse coefficient before drawing any conclusion about the rain gauge.
Q4. Does a logger showing 2.000 mm mean 0.001 mm resolution?
No. The NiuBoL guide notes that extra decimals do not improve physical resolution. Ten tips represent 2 mm on the 0.2 mm configuration, so the meaningful increment is 0.2 mm. Keep the raw tip count, which shows exactly which increments were recorded.
Q5. How many tips per minute is 2 mm/min?
At 0.191 mm per tip it is 2 / 0.191 = 10.5 tips/min, one every 5.7 seconds. At the nominal 0.2 mm it is 10 tips/min, one every 6 seconds. A settle time well below one second is a typical engineering reference value; verify it on the current datasheet.
Q6. Can I use this check for every rainfall type?
No. The injection check covers liquid water delivered slowly near room temperature. Evaporation, splash-out, bucket residue, frozen precipitation and rates above the response ceiling sit outside it. Treat the result as a chain verification, then repeat at several rates if you need a wider operating envelope.
A rain gauge converts water volume into depth through two constants: aperture area and bucket volume. A 200 mm aperture gives 314.16 cm²; a 6 mL bucket makes 0.191 mm per tip; 628 mL should read 20.0 mm. An 18.6 mm reading is a −7.0% deviation to be traced through splash, residue and logger scaling.
Figure 3. Portable Tripods Weather Station Stands.
Verify the coefficient against the physical bucket
A rain gauge reports volume divided by aperture area, not a raw sensor output. A 200 mm aperture and a 6 mL bucket give 0.191 mm per tip, while the 0.2 mm setting implies about 6.28 mL; that small constant difference becomes roughly a 4.7% depth error over a long event. Inject a known 628 mL, compare the 20.0 mm expectation with the logger, and keep raw tip counts next to the scaled depth.
Validating a tipping bucket with a known water volume? Tell us the aperture, bucket size and logger coefficient on our contact page, and we will help you check the per-tip value.
Prev:Greenhouse Climate Control: Verify Each Vent Action
Next:no more
Related recommendations
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related products
Combined air temperature and relative humidity sensor
TDR Soil Moisture & Temperature Sensor for irrigation| RS485 |4-20mA| NBL-S-THR
Soil pH sensor RS485 soil Testing instrument soil ph meter for agriculture |NBL-S-PH
Wind Speed sensor Output Modbus/RS485/Analog/0-5V/4-20mA
0.2mm Tipping Bucket Rain Gauge for Weather & Hydrology | NiuBoL
Pyranometer Solar Radiation Sensor 4-20mA/RS485
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)