— Products —
Consumer hotline +8618073152920 WhatsApp:+8615367865107
Address:Room 102, District D, Houhu Industrial Park, Yuelu District, Changsha City, Hunan Province, China
Water Quality Sensors
NiuBoL NBL-WQ-NHN-4S online ammonia nitrogen sensor adopts ion selective electrode (ISE) technology, with a patented ammonium ion probe and microporous salt bridge design. It is specially designed for wastewater treatment, river monitoring and industrial water quality control. It supports RS485 Modbus RTU protocol, features IP68 protection rating and automatic temperature compensation, solving engineering pain points such as high reagent consumption, high maintenance cost and···
Tel/WhatsApp:+8615367865107
Email:Arvin@niubol.com +Nearly 100 partner companies in more than 75 countries. We are committed to providing high-quality, practical products to meet your needs and help you solve problems. Our products comply with international standards and are certified with ISO, CE and RoHS.Product Details
Product Overview: The NiuBoL NBL-WQ-NHN-4S is an online ammonium ion sensor designed for continuous water quality monitoring. It uses a PVC-membrane ammonium ion-selective electrode (ISE) with automatic Pt1000 temperature compensation. Available measurement ranges are 0–100.00 mg/L and 0–1000.0 mg/L. The sensor features RS485 Modbus RTU communication, optional 4–20 mA output, POM and 316L stainless steel wetted materials, and IP68 protection.
The NBL-WQ-NHN-4S is suitable for evaluating ammonium-related concentrations in wastewater treatment, surface water and compatible industrial water applications. The sensor responds to ammonium ions (NH4+). Measurement performance depends on water chemistry, interfering ions, calibration and the required concentration reporting basis.
Ammonium is an important nitrogen-related water quality parameter in wastewater treatment, environmental monitoring and aquaculture.
The NBL-WQ-NHN-4S uses a PVC-membrane ammonium ion-selective electrode to measure changes in ammonium ion activity in water.
The electrode response is converted into a concentration-related signal using calibration information and temperature compensation.
The sensor incorporates a reference electrode system with a microporous salt bridge. The reference system provides the electrochemical reference required for potentiometric ion measurement.
Unlike reagent-based colorimetric analyzers, an ammonium ion-selective electrode does not consume chemical reagents during routine measurement. However, electrode conditioning, calibration and periodic maintenance are still necessary.
The NBL-WQ-NHN-4S supports RS485 Modbus RTU communication, allowing integration with compatible PLCs, RTUs, data loggers and water quality monitoring equipment.

Ammonium ion-selective electrode: PVC-membrane electrochemical sensing technology.
Two measurement range options: 0–100.00 mg/L and 0–1000.0 mg/L.
Digital communication: RS485 Modbus RTU for integration with industrial monitoring equipment.
Optional analog output: 4–20 mA configuration available when specified.
Automatic temperature compensation: Integrated Pt1000 temperature element.
Reference electrode system: Microporous salt bridge construction.
Two-point calibration: Supports adjustment using appropriate ammonium reference solutions.
Wetted materials: POM and 316L stainless steel.
IP68 protection: Suitable for immersion within the specified operating limits.
3/4-inch NPT mounting: Supports compatible immersion installation arrangements.
Standard cable: 5 meters, with customized lengths available.
Replaceable electrode: Allows electrode replacement when required by condition or measurement performance.
| Parameter | Specification |
|---|---|
| Brand | NiuBoL |
| Model | NBL-WQ-NHN-4S |
| Product Type | Online Ammonium Ion Sensor |
| Measurement Principle | PVC-Membrane Ion-Selective Electrode |
| Measurement Ranges | 0–100.00 / 0–1000.0 mg/L |
| Resolution | 0.01 / 0.1 mg/L |
| Measurement Accuracy | ±10% or ±1 mg/L |
| Temperature Accuracy | ±0.5°C |
| Temperature Compensation | Automatic, Pt1000 |
| Operating Temperature | 0–40°C |
| Operating Pressure | <0.1 MPa |
| Medium pH Range | 4–10 |
| Standard Output | RS485 Modbus RTU |
| Optional Output | 4–20 mA |
| Power Supply | 12–24 VDC |
| Power Consumption | 0.2 W @ 12 V |
| Wetted Materials | POM and 316L Stainless Steel |
| Sensing Membrane | PVC-Based Ammonium-Selective Membrane |
| Installation | Immersion Installation |
| Mounting Thread | 3/4-inch NPT |
| Installation Orientation | Minimum 15° Inclination; No Horizontal or Inverted Mounting |
| Electrical Connector | M16, 5-Core Waterproof Male Connector |
| Standard Cable Length | 5 m; Customizable |
| Calibration | Two-Point Calibration |
| Protection Rating | IP68 |
Measurement selection: The two measurement ranges are alternative product configurations. Select the appropriate range according to the expected ammonium concentration and the measurement requirements of the application.
Output configuration: RS485 Modbus RTU is the standard interface. The optional 4–20 mA configuration should be specified when ordering.
| Measurement Range | Resolution |
|---|---|
| 0–100.00 mg/L | 0.01 mg/L |
| 0–1000.0 mg/L | 0.1 mg/L |
The 0–100 mg/L configuration can be considered where the expected concentration remains within this range.
The 0–1000 mg/L configuration provides a wider measurement range for applications with higher ammonium-related concentrations.
A wider range does not automatically provide better performance at low concentrations.
The specified measurement accuracy is ±10% or ±1 mg/L. For projects requiring formal measurement acceptance, the applicable accuracy criterion should be established for the selected range.
The NBL-WQ-NHN-4S uses potentiometric ion-selective measurement.
The sensing electrode contains a membrane designed to respond preferentially to ammonium ions (NH4+) in water.
When the electrode contacts the water, the activity of ammonium ions contributes to an electrical potential across the sensing membrane.
The sensor measures this potential relative to its reference electrode system.
Using calibration information and temperature compensation, the electrical signal is converted into a concentration-related value.
The electrode response depends on the activity of ammonium ions, which can be influenced by dissolved ion composition and sample conditions.
For reliable measurement, the membrane must be properly conditioned, the electrode must be maintained and the calibration must be suitable for the application.

Ammonium-related measurements can be expressed using different chemical definitions and reporting units.
Understanding these differences is important when comparing online sensor readings with laboratory water quality results.
Ammonium is the positively charged ion NH4+. The NBL-WQ-NHN-4S uses an ammonium-sensitive ion-selective electrode.
Ammonium nitrogen refers to the nitrogen content associated with ammonium ions.
A concentration expressed as mg/L NH4+ differs numerically from the same sample expressed as mg/L NH4-N.
Unionized ammonia is the non-ionized form of ammonia in water.
The relative proportions of NH3 and NH4+ depend on pH, temperature and water chemistry.
NH4+ ⇌ NH3 + H+
Total ammonia nitrogen includes nitrogen associated with both ammonium and unionized ammonia.
An ammonium ISE does not independently measure NH3 and NH4+ as separate channels.
| Term | Meaning |
|---|---|
| NH4+ | Ammonium ion. |
| NH4-N | Nitrogen associated with ammonium ions. |
| NH3 | Unionized ammonia. |
| TAN | Total ammonia nitrogen, including ammonium-N and unionized ammonia-N. |
Important for data integration: When a monitoring project requires results in NH4-N, NH4+ or TAN, specify the required reporting basis during sensor selection and system configuration.
The concentration reporting unit and any conversion method must be consistent with the reference measurements used by the project.
Potassium ions (K+) are an important potential source of interference in ammonium ion-selective electrode measurements.
Potassium can interact with an ammonium-sensitive membrane and contribute to an apparent increase in the measured signal.
The degree of interference depends on the relative concentrations of potassium and ammonium, as well as the sensing membrane characteristics.
For wastewater containing significant potassium concentrations, measurement performance should be evaluated against representative samples.
Other dissolved ions can influence the electrochemical response under certain conditions.
Water composition and ionic strength should be considered when evaluating sensor accuracy.
The specified operating pH range is 4–10.
Changes in pH influence the equilibrium between NH4+ and NH3.
The sensor's direct ammonium-sensitive response should not automatically be interpreted as the total ammonia concentration under all pH conditions.
The integrated Pt1000 element provides automatic temperature compensation for the electrode measurement.
Temperature compensation does not eliminate interference caused by dissolved ions, electrode condition or changes in water composition.
Membrane deposits, aging, inadequate conditioning and reference electrode problems can affect measurement stability.
Regular inspection and appropriate calibration help maintain reliable operation.
The NBL-WQ-NHN-4S can be considered for monitoring ammonium-related concentrations at suitable points within municipal wastewater treatment facilities.
Continuous readings may help operators observe changes in incoming nitrogen-related loading and the conditions associated with biological treatment.
When evaluating nitrification, ammonium measurements can be considered together with dissolved oxygen, pH, temperature and other process parameters.
Automatic aeration or process control requires separate equipment and appropriate control logic.
Industrial wastewater may contain ammonia-related nitrogen compounds at varying concentrations.
The 0–1000 mg/L configuration may be considered where higher concentration monitoring is required and water chemistry is compatible with ion-selective measurement.
For wastewater containing high potassium concentrations or complex chemical mixtures, representative sample testing and reference comparison may be necessary.
Ammonium-related monitoring can provide information about nitrogen-related water quality in rivers, lakes and reservoirs.
Selection should consider the expected concentration, required detection threshold, seasonal water conditions and maintenance requirements.
For very low-concentration applications, confirm that the selected measurement range and accuracy meet the project requirements.
Ammonia-related nitrogen is important in aquaculture because unionized ammonia can affect aquatic organisms.
The NBL-WQ-NHN-4S can provide ammonium-related monitoring data as part of a broader water quality monitoring system.
However, ammonium measurement alone does not determine the concentration of unionized ammonia or its biological effects.
Water pH, temperature and an appropriate ammonia nitrogen reporting basis should also be considered.
The sensor may be considered for selected agricultural drainage and nutrient-rich water monitoring applications.
Where fertilizer-related potassium concentrations are elevated, potassium interference must be considered when evaluating ammonium readings.
The most important selection factors are the required concentration range, measurement reporting basis and actual water chemistry.
| Selection Item | What to Confirm |
|---|---|
| Measurement Range | 0–100 or 0–1000 mg/L. |
| Reporting Unit | NH4+, NH4-N or TAN as required by the project. |
| Concentration Threshold | Minimum concentration or change that must be measured reliably. |
| Water pH | Within 4–10. |
| Water Temperature | Within 0–40°C. |
| Water Chemistry | Potassium and other potentially interfering ions. |
| Installation | Suitable immersion mounting and electrode access. |
| Signal Output | RS485 Modbus RTU or optional 4–20 mA. |
| Reference Testing | Required laboratory method and acceptance criteria. |
For projects requiring a separate measurement of unionized ammonia, total ammonia nitrogen or additional nitrogen compounds, the measurement method and required supporting parameters should be evaluated before selecting the instrument.
The NBL-WQ-NHN-4S is designed for immersion installation using a 3/4-inch NPT threaded connection.
The sensing section should remain exposed to representative water during normal measurement.
The sensor must not be installed horizontally or upside down. Maintain an installation inclination of at least 15 degrees.
Choose a representative monitoring position.
Maintain an installation inclination of at least 15 degrees.
Do not install the sensing electrode horizontally or upside down.
Keep the sensing membrane in contact with water during operation.
Maintain operating pressure below 0.1 MPa.
Keep water temperature within 0–40°C.
Keep sample pH within 4–10.
Protect the sensor from impact and mechanical stress.
Provide access for electrode inspection, cleaning and replacement.
Use waterproof electrical connections and suitable cable protection.
The IP68 rating does not remove the operating pressure or material compatibility requirements.
For pressurized pipelines, a suitable low-pressure sampling arrangement may be required.

The NBL-WQ-NHN-4S operates on 12–24 VDC and supports RS485 Modbus RTU communication.
The standard cable connection is shown below.
| Wire Color | Connection |
|---|---|
| Red | 12–24 VDC Positive |
| Black | Power Ground (GND) |
| Blue | RS485 A |
| Green | RS485 B |
| Yellow | Optional Current Output |
The yellow conductor is associated with the optional analog output. If the current output is not used, leave the conductor disconnected and properly insulated.
Before applying power, verify the wiring against the delivered cable and connector.
All exposed cable joints should be suitably waterproofed and protected against corrosion.

The standard digital output uses Modbus RTU over an RS485 interface.
| Communication Parameter | Default Setting |
|---|---|
| Interface | RS485 |
| Protocol | Modbus RTU |
| Baud Rate | 9600 bps |
| Data Bits | 8 |
| Parity | None |
| Stop Bits | 1 |
| Default Device Address | 6 (0x06) |
| Configurable Address Range | 1–255 |
| Read Function Code | 03 |
| Write Function Code | 06 |
| Register Address | Function |
|---|---|
| 0x0000 | Ammonium-Related Concentration and Temperature, 4 Registers |
| 0x1000 | Lower-Point Calibration |
| 0x1004 | Slope Calibration |
| 0x1010 | Temperature Correction |
| 0x2002 | Device Address Configuration |
| 0x2020 | Restore Factory Calibration Values |
The measurement data block contains four consecutive 16-bit registers: concentration value, concentration decimal scaling, temperature value and temperature decimal scaling.
The following command reads four measurement registers using the default device address 6.
Request Frame:
06 03 00 00 00 04 45 BE
Response Frame:
06 03 08 00 55 00 02 01 18 00 01 B3 5D
The concentration raw value is 0x0055 (85 decimal), and the decimal-place value is 2. The resulting displayed concentration is 0.85 mg/L.
The temperature raw value is 0x0118 (280 decimal), with one decimal place, giving 28.0°C.
This is a communication decoding example. The receiving system must use the correct concentration reporting unit and register scaling.
The NBL-WQ-NHN-4S can provide ammonium-related measurements to compatible PLCs, industrial data loggers and RTUs.
Ammonium Sensor → RS485 Modbus RTU → PLC / Data Logger / RTU → Monitoring Software
For remote monitoring, a separate compatible communication gateway can be used to transmit collected data to a monitoring server.
The sensor supplies measurement data. Automatic aeration, chemical dosing, pumping or treatment control requires additional equipment and suitable control logic.
The NBL-WQ-NHN-4S uses an ion-selective electrode that requires suitable preparation before measurement.
Before first use, immerse the electrode in the appropriate activation solution for approximately 24 hours.
After activation, rinse the electrode with deionized water before testing.
Ensure the sensing membrane is properly conditioned before measurement verification or calibration.
Inspect the sensing membrane for deposits, contamination and physical damage.
Rinse the electrode carefully with distilled or deionized water when cleaning is required.
Avoid hard brushes, abrasive cleaning methods and direct mechanical damage to the PVC membrane.
For storage periods longer than two weeks, keep the sensing electrode dry and protect the sensing element with its protective cap.
Before returning the electrode to service, follow the appropriate conditioning procedure.
Avoid prolonged immersion in distilled water or protein solutions and avoid contact with silicone grease.
The ion-selective sensing electrode is a replaceable component.
If cleaning, conditioning and appropriate calibration cannot restore satisfactory measurement performance, electrode replacement may be necessary.
The NBL-WQ-NHN-4S supports two-point calibration using suitable ammonium reference solutions.
The calibration concentrations and register scaling depend on the selected measurement range.
Place the electrode in the appropriate lower-concentration calibration solution.
Allow approximately five minutes for the reading to stabilize.
Perform calibration using the corresponding Modbus register when adjustment is required.
Place the electrode in the appropriate higher-concentration standard solution.
Allow approximately five minutes for stabilization before performing slope calibration.
The following scaling applies to the respective measurement configurations.
| Measurement Configuration | Calibration Register Scaling |
|---|---|
| 0–100 mg/L | Standard concentration × 100 |
| 0–1000 mg/L | Standard concentration × 10 |
Use the appropriate calibration standard, measurement range and register value for the supplied sensor.
After restoring factory calibration values, the electrode should be recalibrated before normal operation.
| Problem | Recommended Check |
|---|---|
| No RS485 Communication | Check power supply, RS485 wiring, device address, baud rate and cable connections. |
| Unstable Reading | Inspect electrode conditioning, membrane condition, calibration and sample stability. |
| Unexpected High Reading | Check potassium interference, calibration and dissolved ion composition. |
| Unexpected Low Reading | Check electrode activation, sample conditions, selected range and calibration. |
| Difference from Laboratory Results | Confirm concentration reporting units, reference method, pH and temperature. |
| Reading Does Not Stabilize | Inspect sensing membrane condition, electrode storage and reference system. |
A1. It uses an ammonium ion-selective electrode to measure an electrochemical response associated with ammonium ions in water.
A2. The sensor is available with 0–100.00 mg/L and 0–1000.0 mg/L measurement ranges.
A3. No. Total ammonia nitrogen includes nitrogen associated with both ammonium and unionized ammonia. The concentration reporting basis must be considered when comparing measurements.
A4. No. It uses an ammonium-sensitive electrode. The proportion of unionized ammonia depends on water pH and temperature and requires an appropriate calculation or measurement approach.
A5. The specified accuracy is ±10% or ±1 mg/L. Actual performance depends on calibration, water chemistry and the conditions of the measurement.
A6. Yes. Potassium can interfere with ammonium-sensitive electrodes. Applications involving elevated potassium concentrations should be evaluated using representative samples.
A7. It does not consume chemical reagents during routine potentiometric measurement. Activation and calibration solutions are still required for electrode preparation and maintenance.
A8. Yes. The default communication settings are 9600 baud, 8 data bits, no parity, one stop bit and device address 6.
A9. An optional 4–20 mA configuration is available. Specify the required output when ordering.
A10. The sensor operates on 12–24 VDC with a specified power consumption of 0.2 W at 12 V.
A11. The specified water pH range is 4–10.
A12. No. The installation requires an inclination of at least 15 degrees. Horizontal and inverted mounting are not permitted.
A13. The sensor is calibrated before delivery and supports two-point calibration. Cleaning and recalibration should be carried out when required by electrode condition and measurement verification.
A14. It can be considered where the ammonium concentration range, operating conditions and water chemistry are suitable. Interfering ions and reference measurement requirements should be evaluated.
A15. Yes. The RS485 Modbus RTU interface supports connection to compatible industrial data acquisition equipment.
A16. The sensor provides measurement data. Automated aeration or treatment control requires a separate compatible controller and appropriate control equipment.
A17. Provide the required concentration range, reporting unit, water type, pH, temperature, potential interfering ions, output interface, installation requirements, cable length and quantity.
Download the NBL-WQ-NHN-4S user manual for technical specifications, installation instructions, electrical wiring, electrode maintenance, calibration procedures and Modbus RTU register information.
The manual provides the technical information needed to connect the sensor to compatible PLCs, RTUs and data loggers.
NBL-WQ-NHN-4S Online Ammonia Nitrogen Sensor.pdf
For product selection and quotation, please provide the following information:
Measurement requirement: Ammonium ion, ammonium nitrogen or total ammonia nitrogen.
Concentration range: Expected normal and maximum concentrations.
Water application: Municipal wastewater, industrial wastewater, aquaculture or environmental monitoring.
Water chemistry: pH, temperature, potassium concentration and other relevant dissolved ions.
Signal output: RS485 Modbus RTU or optional 4–20 mA.
Installation: Mounting arrangement, immersion conditions and required cable length.
Reference method: Required laboratory comparison or project acceptance procedure.
Order information: Quantity and destination country.
For system integration projects, please also provide the receiving PLC, RTU or data logger model.
Contact NiuBoL for product specifications, technical documentation and quotation.
Explore additional NiuBoL Online Water Quality Sensors for pH, conductivity, dissolved oxygen, turbidity and other water monitoring requirements.
Next:NBL-WQ-NHN-4A Online Ammonium Nitrogen Sensor: For Wastewater Treatment, River and Lake Monitoring
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related recommendations
Online Turbidity Sensor RS485 Modbus | NBL-WQ-TS
Online Ammonia Nitrogen Sensor (Ammonium ISE) | NBL-WQ-NHN
Online Residual Chlorine Sensor (HOCl) RS485 | Water Quality Sensor| NBL-WQ-CL
NBL-WQ-DO Online Fluorescence Dissolved Oxygen Sensor
NBL-WQ-COD Online COD Water Quality Sensor
Water Conductivity EC Sensor RS485 Modbus IP68 | NBL-WQ-EC
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)