— Blogs —
—Products—
Consumer hotline +8618073152920 WhatsApp:+8615367865107
Address:Room 102, District D, Houhu Industrial Park, Yuelu District, Changsha City, Hunan Province, China
Product knowledge
Time:2026-05-27 15:42:44 Popularity:452
Quick Answer: Secondary water supply turbidity monitoring tracks particle-related changes between incoming water, storage and a building's booster or distribution system. Select the sensor from the required low-reading error and sampling arrangement, not its number of decimal places. An immersion probe and a flow-through low-range instrument have different installation needs. Turbidity is an operating indicator; a clear or low-NTU result does not by itself establish microbiological or chemical safety.

In a secondary water supply system, incoming treated water is stored or boosted before reaching users. A turbidity increase can originate upstream, inside storage or in downstream pipework. One outlet probe can detect a change but may not identify where it began. Decide whether the installation needs a permanent inlet comparison, a temporary reference point or paired samples during an investigation.
Tank filling, cleaning, pump starts and changes in demand can disturb particles or introduce bubbles at a sample point. Record those operating states with the reading. A short spike coincident with a pump start needs a different check from a sustained increase present at both inlet and outlet. Neither should be dismissed automatically, but they should not trigger the same diagnosis without investigation.
Secondary water supply monitoring needs a turbidity range and alarm limit appropriate to the applicable water-quality requirement and the purpose of the monitoring point. Define the required reference method, cleaning arrangements and acceptance evidence before selecting an online sensor.
A 0-20 NTU range with 0.01 NTU resolution does not establish 0.01 NTU accuracy. For a project that needs to distinguish 0.10 from 0.20 NTU, display increments are only one part of the decision. Accuracy, repeatability, zero stability, reference uncertainty and the installed sampling arrangement must support that distinction.
The NBL-WQ-TS 0-20 NTU configuration has an accuracy entry of +/-3% or +/-1.5 NTU. The percentage basis and rule for applying those alternatives must be agreed before acceptance. It would be inappropriate to approve a small low-turbidity alarm margin merely because the display shows two decimal places. Ask for demonstrated performance at the intended decision level rather than assuming that the lowest available range is automatically suitable.
For comparison, the NBL-WQ-TSA-4 flow-through configuration covers 0-10.000 NTU with 0.001 NTU resolution and an accuracy expression of +/-5% or +/-0.02 NTU, whichever is greater. Confirm the percentage basis in the accepted specification and test the installed system. These figures distinguish configurations; they do not remove the need to establish method compatibility or suitable performance around a project's alarm threshold.
| Item | NBL-WQ-TS | NBL-WQ-TSA-4 |
|---|---|---|
| Arrangement | Immersion probe | Wall-mounted integrated flow cell |
| Range options | 0-20, 0-100 or 0-1000 NTU | 0-10 NTU |
| Resolution | 0.01 NTU on 20 range; 0.1 NTU on 100/1000 ranges | 0.001 NTU |
| Accuracy expression | 20 range: +/-3% or +/-1.5 NTU; 100: +/-3% or +/-2 NTU; 1000: +/-5% or +/-3 NTU | +/-5% or +/-0.02 NTU, whichever is greater |
| Power supply | 12-24 V DC, +/-10% | 12-24 V DC |
| Consumption at 12 V | Below 0.3 W | 0.2 W |
| Digital output | RS485, Modbus RTU | RS485, Modbus RTU |
| Analog output | Confirm the ordered output version | 4-20 mA optional; confirm assembly |
| Operating temperature; pressure | 0-50 C; below 0.2 MPa | 0-50 C; up to 0.2 MPa |
| Housing; protection | POM and ABS; IP68 | Composite material; IP65 |
| Cable | 5 m standard | 5 m standard |
Confirm cable jacket compatibility and hygienic suitability for the actual installation. IP65 protects the defined flow-through unit from its rated environmental exposure; it does not permit immersion. IP68 on the TS probe does not give its junction box the same rating. Neither protection rating is a potable-water contact approval. Confirm any project-required material documentation separately.

Visual comparison and instrumental light-scattering measurements use different procedures. For continuous monitoring, an online turbidity sensor can provide a signal to a logger or controller. Confirm the optical method, range and reference comparison required by the project rather than comparing instruments only by their displayed unit.
NBL-WQ-TS uses scattered-light measurement. When light enters the water sample, suspended substances scatter the light. The sensor measures scattered-light intensity perpendicular to the incident light, compares it with internal calibration values, and outputs a linearized turbidity value.
Different optical methods and particle populations can produce different results. A reading in NTU should not be converted into suspended solids in mg/L without a suitable site relationship. It does not directly identify odor compounds, dissolved contamination or bacteria. Use turbidity together with the other observations and verification tests required by the water-management plan.
For pump-room water quality, an outlet or bypass that represents delivered water is often more useful than a probe in a stagnant tank corner. Avoid locating the optical region in a persistent bubble plume or against a reflective wall. For the TS immersion arrangement, keep more than 5 cm from the side wall and more than 10 cm from the bottom. Check access for withdrawal and cleaning without contaminating the supply.
A flow-through instrument requires fresh, controlled sample flow and a suitable drain. On the TSA-4 arrangement, use appropriate pressure reduction and keep the discharge path unobstructed. The recommended drain flow is below 300 mL/min; confirm the final hydraulic arrangement rather than applying line pressure directly to the cell. Water hammer and a blocked drain can create conditions unlike steady bench operation.
Sample transport also affects the warning time. In an illustrative calculation, 10 m of tubing with 4 mm internal diameter contains about 0.126 L. At 0.2 L/min, ideal transport through that tube takes about 38 seconds before the cell volume and instrument response are added. A faster logging interval cannot remove that delay. Verify actual flow and replacement behavior after installation.
At 0.2 L/min continuously, an unrecovered sample branch uses 288 L/day. Include this in drainage and utility planning; it is not the instrument's electrical consumption. Any proposed return to the supply needs an approved hygienic design and local review, not an improvised connection. Keep calibration and cleaning liquids from entering the drinking-water circuit.

Turbidity alarm verification should check the measurement, its timestamp and the response separately. For Modbus RTU, set a unique address and compatible serial settings, then confirm register type, scaling and byte order at the PLC, RTU or data logger. A host value of 1.00 must not represent 0.100 because of a misplaced scale factor.
Use a valid known reference to test the whole channel without introducing test liquid into the supply. Simulate loss of sensor communications and, for a flow-through system, loss of sample flow. The display should show an invalid, stale or no-sample condition rather than quietly retaining a plausible value. A probe can communicate successfully while measuring water that has stopped refreshing.
Set persistence and recovery rules around the actual operating purpose. Keep short excursions available for review rather than averaging them away without record. A turbidity event can prompt inspection, a reference check or sampling under the operator's procedure. Automatic interruption of a building's supply needs a separate, approved control strategy; a generic NTU threshold is not enough to define that action.
Acceptance should include suitable low-level references, repeated observations and a comparison using a compatible method. Handle reference solutions and vessels so contamination, settling and bubbles do not dominate the result. A clean-water zero adjustment does not establish accuracy at the alarm level, and a one-point high-turbidity check does not validate the low end.
Routine maintenance includes rinsing the sensor surface with tap water, wiping stubborn dirt with a soft wet cloth, and checking whether the measurement window is dirty. The optical components are sensitive, so the probe should not be hit or disassembled on site.
For the TS calibration arrangement, use appropriate zero and span solutions with the sensing face at least 10 cm above the vessel bottom. Allow the reading to stabilize; the specified procedure uses 3-5 minutes. Treat the TSA flow-cell arrangement according to its assembly-specific procedure rather than dismantling it to imitate an immersion calibration. Record the reference, temperature and result before and after cleaning where useful.
Provide NiuBoL with the normal turbidity, required alarm level, accepted error, reference method, available pressure, sample flow and host inputs. Include the sensor, flow regulation if needed, installation fittings and commissioning in the quotation scope. Selecting a measurement arrangement that fits the pump room is more useful than comparing probe prices while leaving sampling and verification unresolved.

No. Turbidity is an optical particle-related indicator. Chemical and microbiological safety require the appropriate controls and verification. Interpret an NTU reading within the water-management plan rather than as a complete safety certificate.
No. Its accuracy expression, reference method and installed performance must support the required decision level. A 0.01 NTU display step is not a 0.01 NTU accuracy guarantee.
Consider it where a suitable controlled sample branch and drain are available and its low-range performance meets the project requirement. Assess flow loss, bubbles, residence time and service access as part of the measurement.
No. It is a wall-mounted IP65 flow-through assembly, not an IP68 immersion probe. Keep its enclosure and connections within their specified environmental conditions.
There is no single project setting appropriate to every jurisdiction, monitoring location and instrument. Establish the applicable requirement, operating objective and verified measurement uncertainty before setting alarm and response rules.
Check sample flow, pressure changes, bubbles and particle disturbance, then compare a representative reference sample. Retain the event and equipment state; do not automatically delete the spike or assume it proves storage contamination.
Specify range, low-reading error requirement, method, immersion or flow-through arrangement, sample pressure and flow, output, cable route and commissioning scope. Request confirmation of the exact output and wetted-material configuration.
Secondary water supply turbidity monitoring needs low-level measurement performance and a sample point that represents delivered water. Compare the NiuBoL TS and TSA-4 arrangements by error, hydraulics and service access, then verify the whole channel before relying on its alarms. Preserve separate chemical and microbiological checks required for the water supply.
Operational context: WHO sanitary inspection and maintenance guidance for piped distribution.
NBL-WQ-CL Water Quality Sensor Online Residual Chlorine Sensor.pdf
NBL-WQ-DO Online Fluorescence Dissolved Oxygen Sensor.pdf
NBL-WQ-NHN Ammonia Nitrogen Water Quality Sensor.pdf
NBL-WQ-COD Online Water Quality COD Sensor.pdf
NBL-WQ-PH Online pH Water Quality Sensor.pdf
NBL-WQ-EC water quality conductivity sensor.pdf
Prev:NBL-WQ-TSS Applications: Settling Outlet vs Concentrated Sludge
Next:Industrial Water Quality Requirements by Cooling, Boiler and Process Use
Related recommendations
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related products
Combined air temperature and relative humidity sensor
TDR Soil Moisture & Temperature Sensor for irrigation| RS485 |4-20mA| NBL-S-THR
Soil pH sensor RS485 soil Testing instrument soil ph meter for agriculture |NBL-S-PH
Wind Speed sensor Output Modbus/RS485/Analog/0-5V/4-20mA
0.2mm Tipping Bucket Rain Gauge for Weather & Hydrology | NiuBoL
Pyranometer Solar Radiation Sensor 4-20mA/RS485
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)