— Blogs —
—Products—
Consumer hotline +8618073152920 WhatsApp:+8615367865107
Address:Room 102, District D, Houhu Industrial Park, Yuelu District, Changsha City, Hunan Province, China
Product knowledge
Time:2026-09-12 12:00:00 Popularity:198
Quick Answer: Self-recording rain gauge installation must verify the rainfall measurement and its time record independently. Level the collector and check water delivery through the instrument, then prove the chart clock or electronic logger, interval boundaries and restart behavior. A correct event total can still have the wrong timing, and a live display does not prove that the historical record is complete.

A mechanical chart recorder preserves a trace on paper. A tipping-bucket gauge supplies pulses to a recorder, while a configured digital transmitter may expose a rainfall value to a data logger. A gauge body without the associated recording equipment is not automatically a complete historical monitoring system.
Start the installation schedule with the actual measurement and recording chain. For a chart instrument, identify the clock movement, chart scale and change procedure. For an electronic system, identify the pulse counter or serial input, storage interval, clock, memory and export method. Remote telemetry is a separate capability, not the meaning of self-recording itself.
Where the project is about an existing gauge fault rather than a new record, use the separate tipping-bucket maintenance guide. This procedure focuses on whether a new installation creates a complete, correctly timed rainfall record.
Choose an exposure suited to the observation method, with a stable base and clear collector opening. Avoid sprinkler spray, roof runoff, splash and foreseeable obstruction by vegetation. Record collector height and surrounding features rather than treating one height as a universal requirement for every gauge and observing network.
Level the instrument using its adjustment arrangement, tighten the fixing points and recheck. Transport restraints must be removed where fitted. Confirm that the drain path remains unobstructed after the outer housing is installed. Installation photographs should show the completed condition, not only an open mechanism before it is secured.

The exposed tipping mechanism above is a pulse-generating arrangement, not a paper-chart siphon recorder. In either case, keep the measurement path and service access clear. Do not move or adjust the collector after the baseline test without recording the change.
For an electronic option, the NiuBoL NBL-W-RS 0.2 mm configuration has the following relevant specifications. They do not describe the measuring scale or clock of a mechanical chart instrument.
| Item | NBL-W-RS 0.2 mm configuration |
|---|---|
| Rainfall increment | 0.2 mm per tip; nominal 6.28 mL tipping volume |
| Collector diameter | 200 +/-0.6 mm |
| Intensity range | 0-4 mm/min |
| Specified accuracy condition | +/-4% for the stated indoor test at 2 mm/min |
| Output versions | Reed-switch pulse, voltage or RS485; identify the actual version |
| Supply versions | 5 V DC or 12-24 V DC according to configuration |
| Working temperature | 0-50 deg C; not an unheated snow measurement rating |
| Standard cable | 5 m; confirm extensions, sheath and connection protection for the site |
Confirm the delivered enclosure and connector protection separately, and budget any powered transmitter, logger and modem independently. Do not apply a station-wide wattage or IP rating from the appearance of the stainless-steel collector.
Test the receiving increment before collecting natural rain. With an illustrative 0.2 mm pulse setup, 15 accepted tips should add 3.0 mm. This is a count-and-scaling check, not a quantitative hydraulic calibration. A controlled water test and its reference method remain separate.
Record a reference time and the logger or chart time at installation. Repeat the comparison after an appropriate operating period. A chart drum can turn while its timing is wrong, and a digital device can upload successfully while using the wrong time zone.
For a paper chart, align it correctly on the drum and mark the actual start time. Keep chart identity, date and any clock correction with the original record. A replacement chart should carry the end and start marks needed to distinguish an overlap from a gap.
For electronic logging, define whether the stored timestamp identifies an interval start, interval end or individual tip. Record the daily boundary and how the system treats clock synchronization. Changing the time zone in a dashboard should not silently shift the raw observation history.
The station below illustrates a wider weather installation. Its shared clock and logger may support several channels, but the rain record still needs its own accumulation and reset tests. A time setting that appears correct in an air-temperature graph can be wrong for daily rainfall totals.

If a transmitter returns cumulative values of 42.0 and 43.4 mm, the intervening increment is 1.4 mm. Repeatedly adding cumulative readings creates false rainfall. If the instrument resets or rolls over, detect that event explicitly rather than accepting a negative increment.
Test the interval boundary deliberately. For example, make accepted test tips immediately before and after a ten-minute boundary, then verify that the total across both records equals the total test count. Do not impose the example interval on every project; use the actual configured period.
A daily total and an event total may have different boundaries. A storm crossing midnight should not lose its continuity merely because the daily counter changed. Keep the observation increments or tip times needed to reconstruct both views.
Disconnect the uplink while leaving the field logger operating. Confirm what observations remain local, whether the memory has enough capacity for the required outage, and how stored data are recovered. Telemetry failure should be distinguishable from a sensor or counting failure.
On reconnection, confirm that the receiver does not duplicate previously transmitted records or assign old rain to the upload time. A restored connection is not proof that the missing historical period has been restored. Check the complete expected sequence of stored intervals.
Size storage from the actual record format. At five-minute intervals, one channel produces 288 records per day and 25920 records in 90 days. If a design assumes 32 bytes per record, that is 829440 bytes, about 0.83 MB, before headers, diagnostics and other weather channels. This is a capacity example, not a specification for a NiuBoL logger. Individual tip timestamps can require a different budget because record count then depends on rainfall.
Test what happens when available storage is exhausted: an instrument may overwrite the oldest data, stop recording or require an operator action. Obtain the selected logger's actual behavior and verify export or backfill before relying on an unattended storage period. Keep observation time distinct from the time at which the server receives the recovered record.
A solar-powered field station such as the installation below also needs a controlled power-recovery test. Determine whether the logger retains the clock, count and configuration after interruption. Identify an unobserved period as missing rather than zero rainfall.

A soil-temperature or moisture probe can help study the field response to rain, but it cannot verify rainfall accumulation. Its units, depth and response time differ from a rain gauge. A dry soil trace after a small shower is not by itself evidence that the rain record is wrong.

Likewise, agreement between two graphs does not establish calibration. Accept the water-collection response, count or chart scale and time record through their own tests. Mark all artificial test water and maintenance times before enabling operational reports.
An electronic historical record needs suitable recording equipment, whether built in or separate. A mechanical chart instrument records differently and does not acquire digital telemetry merely by being called automatic.
Yes. Rain may be assigned to the wrong short interval or day even when a longer cumulative sum agrees. Test clock accuracy and accumulation boundaries separately from total volume.
No. It can verify movement and the count path. Quantitative rainfall calibration needs a suitable controlled water delivery and reference procedure under the specified conditions.
Use the configured digital transmitter's actual Modbus RTU settings and data definition. A reed-switch contact is not an RS485 device. Verify whether the returned quantity is cumulative rainfall, an interval amount or another defined value.
The required rainfall observations, original times, quality flags and reset history should remain available according to the purchased logger capability. State the required storage duration and test recovery before acceptance.
List the gauge, recording method, mounting, cable, input hardware, power and remote communication. Include clock and record-verification work separately from a gauge-only supply, with the actual reporting intervals and storage requirement.
Accept the rainfall and time record as separate parts of the installation. Correct scaling, preserved interval boundaries and tested outage recovery make the history usable after the first live display is gone. Send NiuBoL the intended recording method, logger interface and required storage interval to define the appropriate gauge and acquisition package.
Prev:Automatic Soil Water Monitoring System: Principle and Maintenance
Next:NBL-W-HX1 Siphon Rain Gauge: Structure, Principle and Project Use
Related recommendations
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related products
Combined air temperature and relative humidity sensor
TDR Soil Moisture & Temperature Sensor for irrigation| RS485 |4-20mA| NBL-S-THR
Soil pH sensor RS485 soil Testing instrument soil ph meter for agriculture |NBL-S-PH
Wind Speed sensor Output Modbus/RS485/Analog/0-5V/4-20mA
Tipping bucket rain gauge for weather monitoring auto rainfall sensor RS485/Outdoor/stainless steel
Pyranometer Solar Radiation Sensor 4-20mA/RS485
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)