— Blogs —
—Products—
Consumer hotline +8618073152920 WhatsApp:+8615367865107
Address:Room 102, District D, Houhu Industrial Park, Yuelu District, Changsha City, Hunan Province, China
Product knowledge
Time:2024-09-24 16:00:04 Popularity:1946
Weather station maintenance should end with a measurement check, not just a clean appearance. Record the pre-service condition, identify observations affected by cleaning or testing, and verify each disturbed channel before normal reporting resumes. Rain-gauge test tips, a moved wind reference or a wet air-sensor shield can change records even when the station continues communicating.

A service visit can remove a genuine obstruction and also introduce a new measurement problem. Cleaning the rain collector may generate pulses. Replacing a wind bracket may alter its direction reference. An enclosure inspection may disturb a cable without immediately stopping communication. These effects are easier to recognize when the original condition and service times are recorded.
Before work, note the affected channels, current values, timestamps and visible condition. Photograph orientation or level references where relevant. Use a safe service window and the approved access and isolation procedures. Do not approach exposed station equipment during a thunderstorm or open damaged electrical equipment merely to complete a cleaning checklist.
Mark the service interval in the reporting process so a later analyst can distinguish environmental observations from technician activity. Retain the original data and describe any exclusions or corrections. Silently deleting unusual values can erase the evidence needed to understand the maintenance.

Inspect the collector opening, funnel, outlet and accessible mechanism using the procedure for the fitted gauge. Do not scrape, bend or adjust the bucket assembly simply to improve its appearance. The cleaning method for the collector housing is not automatically suitable for the measurement mechanism.
For an NBL-W-RS with 0.2 mm resolution, three deliberate test tips would represent 0.6 mm in a correctly configured counting system. This is an illustrative conversion, not a calibration procedure. Check where those increments appear: local input, logger interval and platform daily total.
A functional count check does not establish the gauge's rainfall accuracy. Its stated +/-4% accuracy relates to an indoor static test at 2 mm/min. Any calibration requires the appropriate controlled method. Do not pour an unmeasured bottle of water into the funnel and call agreement with an expected count a calibration certificate.
| After-service observation | Check before accepting the record | Do not do this |
|---|---|---|
| A rainfall increment appears during cleaning | Service times, deliberate tips and the receiving system's accumulation rules. | Report it as natural rainfall without review. |
| Test tips occur but the logger does not count | Selected output version, connection and the actual input. | Assume a clean funnel proves the full signal path works. |
| The gauge was tilted or remounted | Level and exposure after reassembly. | Return it to service based only on a plausible total. |
If real rain occurs during the work, test increments and natural precipitation may overlap. Do not assume the natural amount can always be reconstructed. Flag the affected period and report its limitation rather than inventing a corrected event total.

For NBL-W-SS and NBL-W-DS wind instruments, check unobstructed movement and the final mounting condition. Both models have IP65 ingress protection. Do not add oil or protective coatings unless the specific service procedure authorizes them. Moving the cups by hand is not a calibrated wind-speed test.
If the vane or crossarm was removed, restore the documented direction reference and verify the receiving channel. A plausible direction that has shifted with the bracket can be harder to detect than a communications failure. Compare against the intended orientation, not an approximate photograph taken from a different viewpoint.
The NBL-W-LBTH shield should retain its ventilation paths. Cleaning must not force water or debris into the sensing area. If moisture or handling has affected the observation, allow the instrument to recover as appropriate and review the subsequent trend. Do not alter an offset solely to force immediate agreement with an instrument in different exposure.
| Component | Relevant measurement detail | Post-maintenance check |
|---|---|---|
| NBL-W-LBTH | Air temperature accuracy +/-0.5 C; humidity accuracy +/-5% RH; selected RS485 Modbus RTU or analog output. | Open louvers, representative airflow, correct channel units and fresh records. |
| NBL-W-DS | Wind direction 0-360 degrees; accuracy +/-3 degrees; IP65. | Restored reference and free vane movement, without confusing a functional check with calibration. |
| NBL-W-HPRS | Solar radiation range 0-2000 W/m2; selected output options. | Clean, undamaged sensing surface, intended plane and no service cover or new shadow. |
A solar-radiation sensor must receive the radiation it is intended to measure. Do not leave a sunshade, cloth or protective film over it as a general overheating precaution. Clean the particular sensing surface using an approved method, then check its level or plane, cable condition and surrounding shadows. A zero reading at night is not a complete calibration check.
A permanently buried soil probe should not be pulled out and wiped on every visit. Its soil contact, depth and position are part of the measurement. If removal is necessary, record it as an installation change and verify the restored contact. A reading after reinsertion should not automatically be compared with the old trend as though nothing changed.

Compare the actual sensor or receiver channel with the record arriving at the destination. Confirm time, units, scale and freshness. A platform may continue displaying its last valid value after a cable is disconnected, so an unchanged screen is not enough to demonstrate successful restoration.
For RS485 Modbus RTU versions, check address and serial settings after any replacement or reset. For an analog or pulse version, verify the actual input configuration. Do not reset cumulative totals or change register scaling merely to make a graph appear continuous after maintenance.
Software changes should follow an approved compatibility and recovery procedure. Installing the newest firmware during an unrelated cleaning visit can add another cause of changed behavior. Keep configuration changes identifiable and confirm the data path afterward.

A blocked collector after a short interval requires a different response from a clean collector after a long one. Consider vegetation, insects, dust, irrigation and seasonal exposure when planning the next inspection. Do not impose one annual calibration or cleaning interval on every instrument without the relevant maintenance requirements and measurement objective.
Close the service record with the work performed, channels affected, test-generated data, final orientation or level checks and any unresolved limitation. For exact cleaning methods, consult the weather station maintenance guide. NiuBoL support enquiries should identify the installed model, output version and the symptom before and after the work.

They should not be silently reported as natural rainfall. Identify the service interval and test increments, then apply the agreed reporting treatment while retaining the original record. Overlapping real rain may prevent a reliable reconstruction.
No. It is a useful mechanical check, not a controlled accuracy assessment. Calibration needs an appropriate reference and method. Also verify the restored direction reference if the vane or bracket was moved.
No, not as a general maintenance measure. It must observe the required radiation in the intended plane. Remove temporary covers and check that service equipment or changed brackets have not introduced a shadow.
No. A display can retain an old value. Verify acquisition timestamps, valid communication and the destination record. Check each disturbed channel rather than judging the station by one updating field.
No. Routine removal disturbs the measurement location and contact. Remove only when the specific task requires it, record the disturbance and verify reinstallation before interpreting the new readings.
No universal interval suits every sensor and duty. Follow applicable instrument requirements and the project's quality plan, considering exposure, drift evidence, past faults and the consequences of an incorrect observation.
Weather station maintenance should preserve the meaning of the observations. Check what cleaning, handling and testing changed, verify the restored data path, and identify service-generated records. A clean station is ready for normal reporting only when the relevant measurements and their limitations have been reviewed.
Prev:Micro Weather Station Purchase: Clarify After-Sales Support and Spares
Next:Weather Station Surge Protection: Match Power and RS485 Circuits
Related recommendations
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related products
Combined air temperature and relative humidity sensor
TDR Soil Moisture & Temperature Sensor for irrigation| RS485 |4-20mA| NBL-S-THR
Soil pH sensor RS485 soil Testing instrument soil ph meter for agriculture |NBL-S-PH
Wind Speed sensor Output Modbus/RS485/Analog/0-5V/4-20mA
Tipping bucket rain gauge for weather monitoring auto rainfall sensor RS485/Outdoor/stainless steel
Pyranometer Solar Radiation Sensor 4-20mA/RS485
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)