Quick Answer: A full spectrum water quality sensor can provide several optical and concentration-equivalent trends through one digital connection. Select it by the required outputs and their demonstrated relationship to the site's water, not the number of parameter names on a screen. For NiuBoL FUV-200 and FUV-408, confirm the exact configuration, full-scale error and independent field comparisons before using a channel for an operating decision.

The sensor records an optical response and applies parameter-specific calibration or correlation models. COD, TOC and BOD outputs therefore need a defined reference basis. They are not automatically equivalent to digestion-based COD, a separate carbon analysis or a five-day biological test in every wastewater.
Several channels rising together may be useful evidence of a sample change, but they are not independent confirmations of the same pollution event. A common change in the optical window or matrix can affect multiple outputs. Keep the measured signal, derived concentration and laboratory reference method distinct when specifying a river warning station or industrial wastewater point.
The NBL-WQ-FUV-200 covers 200-800 nm and lists five output parameters. The FUV-408 lists additional nutrient-related outputs. Do not assume those extra outputs are present on FUV-200 or that every listed equivalent has been validated in the intended stream.
| Output or condition | FUV-200 | FUV-408 |
|---|---|---|
| COD equivalent | 0-200 mg/L equiv. KHP | 0-200 mg/L equiv. KHP |
| TOC; BOD | Each 0-150 mg/L | Each 0-150 mg/L |
| Chromaticity; turbidity | 0-500 Hazen; 0-400 NTU | 0-500 Hazen; 0-400 NTU |
| Additional listed outputs | No nutrient channel assumed | TP 0-15; TN 0-100; NH3-N 0-80; nitrate 0-15; nitrite 0-10 mg/L; confirm bases and applicability |
| Accuracy | +/-5% F.S. for each listed parameter | +/-5% F.S. for each listed parameter |
| Common resolution | COD/TOC/BOD 0.1 mg/L; chromaticity 0.1 Hazen; turbidity 0.1 NTU | Same for these five outputs |
| Water conditions | 0-45 degrees C; pressure below 0.1 MPa | 0-45 degrees C; pressure below 0.1 MPa |
| Supply; listed power | 12-24 VDC; 0.2 W at 12 V | 12-24 VDC; 0.2 W at 12 V |
| Interface; construction | RS485 Modbus RTU; 316L; IP68; standard 5 m cable | RS485 Modbus RTU; 316L; IP68; standard 5 m cable |
| Cleaning configuration | Brush option to be confirmed | Listed with mechanical brush; confirm ordered assembly |
Confirm cable jacket, connector protection and cleaning loads separately. The listed low power figure is not sufficient to size a complete station with logger, telemetry and powered cleaning. Do not add a 4-20 mA requirement without confirming a suitable transmitter or conversion arrangement; these FUV configurations specify RS485 output.

At a 200 mg/L COD span, 5% of full scale is 10 mg/L. The corresponding terms are 7.5 mg/L for a 150 mg/L TOC span and 20 NTU for a 400 NTU turbidity span. These are specification calculations, not the complete uncertainty of a field installation.
A project aiming to distinguish a small COD change in low-organic river water must consider that absolute term. The 0.1 mg/L display increment does not establish the required capability. Obtain the applicable reference conditions and evaluate the actual decision band before selecting the instrument.
Do not transfer a COD correction factor to TOC or BOD simply because all use mg/L. Spectral baseline calibration in deionized water checks a different issue from site-specific concentration agreement. A stable baseline cannot establish that a changed industrial recipe follows the previous COD relationship.

The separate COD probe pictured above is another measurement configuration, not an automatic independent reference. Likewise, the turbidity probe below provides a separate optical observation only if selected and validated for that duty. Adding it does not automatically correct the FUV concentration models.
FUV turbidity compensation should be evaluated within the specified 0-400 NTU band and representative particle conditions. Strong absorption, deposits or bubbles can still undermine a result. When several FUV outputs change after cleaning, inspect window condition before interpreting the event as simultaneous chemical changes.
For these FUV installations, arrange horizontal fixation with the optical region facing refreshed flow, at least 30 cm below the lowest water level. The FUV-408 installation guidance also recommends a depth no greater than 2 m; confirm the supplied drawing and site arrangement. Avoid bubble plumes and dead zones, and provide safe retrieval for window inspection.
At the PLC, RTU or data logger, verify each register, unit, scaling and channel label separately. Distinguish maintenance and lost communication from measured zero. Record site-model changes so archived observations remain interpretable rather than silently mixing different conversion relationships.

No. Specify the actual outputs of the ordered model and the reference basis for each. Do not infer pH, dissolved oxygen or an unlisted nutrient channel from the term full spectrum.
Not automatically. Validate the optical relationship and use the analytical method accepted for the intended reporting or process duty.
No. They can share optical and matrix influences. Use an appropriate independent comparison when the operating decision requires confirmation.
No. A clean-water baseline and a sample-specific concentration relationship are different checks. Review actual-water agreement when composition changes materially.
It produces an absolute error term that may be significant at low concentration. Compare that term and field uncertainty with the change the station must distinguish.
Send required outputs, concentration bands, decision margins, reference methods and representative water information. Add turbidity, color, installation and cleaning conditions, receiver and power constraints so the trial and quotation cover the intended duty.

A full spectrum water quality sensor is useful when its required outputs are demonstrably suited to the sample and decision. Confirm the FUV configuration, check full-scale error and validate each concentration relationship independently. A multiparameter probe assembly is a different architecture; select its actual modules rather than assuming it supplies the same optical outputs.
Prev:MLSS Sensor Field Calibration: Paired Samples for Aeration and Return Sludge
Next:Combination pH Electrode: Reference Junction, Transmitter and Storage
Related recommendations
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related products
Combined air temperature and relative humidity sensor
TDR Soil Moisture & Temperature Sensor for irrigation| RS485 |4-20mA| NBL-S-THR
Soil pH sensor RS485 soil Testing instrument soil ph meter for agriculture |NBL-S-PH
Wind Speed sensor Output Modbus/RS485/Analog/0-5V/4-20mA
Tipping bucket rain gauge for weather monitoring auto rainfall sensor RS485/Outdoor/stainless steel
Pyranometer Solar Radiation Sensor 4-20mA/RS485
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)