— Blogs —
—Products—
Consumer hotline +8618073152920 WhatsApp:+8615367865107
Address:Room 102, District D, Houhu Industrial Park, Yuelu District, Changsha City, Hunan Province, China
Product knowledge
Time:2026-06-13 10:04:40 Popularity:464
Quick Answer: Turbidity and suspended solids sensor selection starts with the decision at the treatment point. Choose an NTU measurement for a defined clarity or particle-breakthrough task; choose a validated mg/L solids measurement for concentration or solids loading. Then check the operating range, sample matrix, installation and acceptance tolerance. A wide-range optical probe is not automatically suitable for low-level effluent or concentrated return sludge.

A wastewater specification that asks for a "particle sensor" leaves an important choice unresolved. The outlet operator may need an early warning of carryover, while the process engineer needs solids concentration for a mass balance. The electrical contractor needs a signal list with units and scaling. Agree on that task before comparing probe prices.
Turbidity measures an optical scattering response, commonly expressed in NTU. An optical TSS sensor estimates suspended-solids concentration through a calibration relationship, usually expressed in mg/L. It does not weigh particles continuously. For the underlying distinction and conversion limits, see TSS versus turbidity measurement. Here the practical question is which instrument belongs at each treatment point and how to accept it.
| Treatment point | Decision to support | Selection consequence |
|---|---|---|
| Raw industrial inlet | Detect a change in incoming particles or estimate incoming solids load | Check peaks, oil, color and particle changes; do not assume one calibration covers all production recipes. |
| Clarifier outlet | Recognize carryover or quantify effluent solids | NTU can support a clarity alarm; mg/L requires a validated concentration relationship near the decision value. |
| Filter outlet | Identify a small departure from normal clear water | Evaluate low-level error and sample conditioning, not only the smallest displayed increment. |
| Aeration basin | Follow mixed-liquor concentration | Check the actual MLSS range and aeration effects; a 2000 mg/L upper limit may be insufficient. |
| Return or concentrated sludge | Support sludge transfer, inventory or loading calculations | Measure expected concentrations first and select a range intended for that duty. |
A turbidity alarm in clarifier effluent monitoring may be useful even when no defensible NTU-to-mg/L conversion exists. Keep it labelled as turbidity. Conversely, when a mass balance or a concentration limit is the task, an attractive NTU trend is not a substitute for concentration verification.
Collect normal, peak and upset values in the required unit. Record whether the values came from whole-water samples, filtered samples, an existing online instrument or an estimate. An assumed conversion from NTU is not a measured TSS peak. Include startup, sludge wasting, backwash and production changes when those conditions will reach the probe.
Suppose a basin normally contains 3200 mg/L of suspended solids and the selected model stops at 2000 mg/L. That probe is out of range during ordinary operation. A configurable PLC display cannot increase the physical measuring range. Equally, selecting a very high upper range does not prove sufficient low-level performance at a final effluent point.
For sludge concentration monitoring, specify whether the result is mg/L, g/L or percent solids and define any conversion basis. For example, 2000 mg/L is 2 g/L; it is not 2% solids. Converting a liquid mass concentration to mass percent also requires a consistent density basis. Keep engineering units explicit in the controller and operating screen.

The comparison below uses the NBL-WQ-TS immersion turbidity configuration and NBL-WQ-TSS-4A suspended-solids configuration. The stainless steel NBL-WQ-TSS-4S is a different option with its own range and material selection. Housing appearance does not establish measurement performance.
| Parameter | NBL-WQ-TS immersion turbidity | NBL-WQ-TSS-4A suspended solids |
|---|---|---|
| Measurement and range | 90-degree infrared scattering; 0-20, 0-100 or 0-1000 NTU options | Scattered-light concentration estimate; 0-2000 mg/L |
| Stated accuracy | 0-20 NTU: +/-3% or 1.5 NTU; 0-100 NTU: +/-3% or 2 NTU; 0-1000 NTU: +/-5% or 3 NTU | +/-5%, dependent on sludge homogeneity; temperature +/-0.3 degrees C |
| Resolution | 0.01 NTU for 0-20; 0.1 NTU for other listed ranges | Obtain the concentration resolution in mg/L for the ordered version |
| Operating temperature and pressure | 0-50 degrees C; below 0.2 MPa | 0-50 degrees C; below 0.2 MPa |
| Supply and power | 12-24 VDC +/-10%; below 0.3 W at 12 V | 12-24 VDC; 0.2 W at 12 V |
| Output | RS485 Modbus RTU | RS485 Modbus RTU; 4-20 mA configuration to be confirmed in the order |
| Installation and protection | Immersion; 3/4 NPT; IP68 | Immersion; 3/4 NPT; IP68 |
| Housing and cable | POM/ABS configuration; 5 m cable | POM/ABS configuration; 5 m cable |
The purchase specification needs the percentage basis and the rule for combining percentage and absolute turbidity errors. Neither a percentage of reading nor a percentage of full scale can be assumed without that definition. For TSS-4A, require a concentration resolution stated in mg/L; an NTU increment is not a solids-concentration resolution. Confirm cable jacket compatibility separately when oil, cleaners or industrial chemicals are present.
A display increment of 0.01 NTU does not prove accuracy at 0.01 NTU. If the task is detecting a small change in filtered water, compare the required alarm band with the specified error, repeatability and performance in representative water. A dedicated low-turbidity flow-through instrument may deserve evaluation instead of an immersion probe.
An optical TSS sensor acceptance trial should cover the liquid the instrument will actually see. Collect paired online readings and reference samples from a representative, safely accessible point. Match timestamps after allowing for sample transport. Keep mixing adequate for representativeness without creating an artificial bubble problem or breaking flocs in a way that differs from operation.
For illustration, a reference result of 20 mg/L paired with a sensor result of 30 mg/L is a +10 mg/L difference, or +50% relative to that reference. Whether that is acceptable depends on the specified task and comparison uncertainty; a visually smooth trend does not answer the question. At very low reference values, relative percentages can become misleading, so define an absolute tolerance where appropriate.
A comparison based only on stable normal operation does not establish behavior during an oily batch, a coagulant change or a solids washout. If the relationship changes with particle type, either restrict the validated operating conditions or use separate validated calibrations where the instrument permits them. Do not hide the change by fitting one broad conversion line.

For the listed immersion configurations, keep the sensing area more than 5 cm from a side wall and more than 10 cm from the bottom. Provide a rigid, retrievable support that does not carry probe weight through the cable. Choose a point with representative flow, away from settled pockets and direct bubble plumes. Confirm actual clearance with the measuring end in its installed position.
A small sample chamber can improve access but can also change the sample. Settling in a long hose removes particles; pressure reduction can release bubbles; a screen may remove the very solids the project intends to measure. Identify which parts of the sample train are allowed and validate the whole arrangement, not just a probe in a beaker.
For RS485 Modbus RTU integration, verify address, baud rate, register type, scale factor and unit on a single instrument before connecting the full bus. The PLC should distinguish an invalid measurement or timeout from a valid zero. Keep cleaning and calibration states separate from a process carryover alarm. Any analogue output must have a confirmed range assignment and failure indication.
An illustrative 4-20 mA signal scaled to 0-2000 mg/L gives 1000 mg/L at 12 mA: (12-4)/16 times 2000. This checks cabinet scaling only, not optical accuracy. A correctly decoded current can still come from an unrepresentative location, dirty window or unsuitable calibration.
Request separate prices for the selected sensor, holder, controller or gateway, required cable, sample arrangement and cleaning equipment. Identify which items are included. A self-cleaning version and a manual-cleaning probe should not be compared as identical packages, and an IP68 label does not establish an automatic wiper or chemical compatibility.
Estimate maintenance from an observed fouling trial. Record the clean reading, the reading before cleaning and the reading after cleaning while accounting for process changes. The time before fouling creates an unacceptable error determines a practical inspection interval. There is no reliable universal monthly cleaning schedule for all wastewater particle sensors.
Send NiuBoL the treatment point, expected concentration distribution, reference method, operating pressure and temperature, bubble exposure and control decision. Include the required cabinet signal and trial acceptance criteria. That information supports a choice between turbidity, suspended solids or a higher-range sludge instrument without treating them as interchangeable probes.

Use turbidity for a defined clarity or carryover trend. Use a site-validated solids measurement when the task requires mg/L. Evaluate both near the expected outlet concentration and alarm value; neither unit is automatically preferable at every clarifier.
Only if its range and validated performance cover the actual stream. Measure normal and peak concentrations first. A return-sludge line above 2000 mg/L needs a different measuring range, not a larger number entered in the display.
No. It does not remove bubbles, clean the window or correct an arbitrary change in particle size, color or composition. Those effects need installation control, maintenance and sample comparisons.
No. A strong correlation can still include a substantial offset or scale error. Check independent validation samples and errors around the process decision, using the reference method and tolerance agreed for the project.
Not when filtration removes the suspended material being evaluated. Define the reference fraction and method first. A dissolved-fraction sample cannot validate a whole-water suspended-solids result.
Do not assume simultaneous outputs across a family. Specify the exact ordered configuration, wiring, scaling and controller input. The digital TS configuration in the table should not be assigned an analogue output without confirmation.
Compare the same installed scope, including the holder, cleaning mechanism, controller, consumables and access requirements. Include site-trial results when estimating service labor; a lower purchase price alone does not establish the lower maintenance cost.
A factory test does not establish method approval or performance in every wastewater matrix. Confirm the applicable reporting requirements and complete the agreed site verification before using the result for that purpose.
Turbidity and suspended solids sensor selection is successful when the unit, range and installed measurement answer the treatment-point decision. Screen out unsuitable ranges first, resolve the accuracy expression, and accept the optical response against representative samples. NiuBoL can scope the appropriate particle sensor once those process requirements are clear.
Prev:Conductivity Factors in Water Quality: Temperature, Mixing and Fouling
Next:Turbidity Testing and Chloride: Separate Particle and Dissolved-Ion Checks
Related recommendations
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related products
Combined air temperature and relative humidity sensor
TDR Soil Moisture & Temperature Sensor for irrigation| RS485 |4-20mA| NBL-S-THR
Soil pH sensor RS485 soil Testing instrument soil ph meter for agriculture |NBL-S-PH
Wind Speed sensor Output Modbus/RS485/Analog/0-5V/4-20mA
Tipping bucket rain gauge for weather monitoring auto rainfall sensor RS485/Outdoor/stainless steel
Pyranometer Solar Radiation Sensor 4-20mA/RS485
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)