— Blogs —
—Products—
Consumer hotline +8618073152920 WhatsApp:+8615367865107
Address:Room 102, District D, Houhu Industrial Park, Yuelu District, Changsha City, Hunan Province, China
Product knowledge
Time:2024-11-19 15:38:24 Popularity:2491
Aquaculture equipment sensors need different checks for different measurements. Cleaning removes deposits, verification compares a reading with an appropriate reference, and calibration changes the instrument's response. None of those actions alone proves that an aerator, pump or dosing system operates correctly. For NiuBoL online sensors, use the procedure for the actual model, confirm the reference quantity and separate service readings from production-water observations.

For the NBL-WQ-PH, identify the configured buffer set and use suitable fresh buffers at a known temperature. Its procedure uses pH 6.86 for the central point and pH 4.00 or 9.18 for the applicable slope check. Those nominal values are not interchangeable with every other commercial buffer set or with the solution's value at another temperature.
Inspect and clean the electrode as appropriate, allow the reading to stabilize, and compare before writing new calibration values. Retain the original check result. A probe that is fouled, dry or installed incorrectly should not be made to appear correct by repeatedly adjusting an offset.

The NBL-WQ-TS turbidity sensor uses scattered light. Clean the optical window, avoid bubbles and use the specified zero and standard solutions. Its calibration arrangement keeps the sensor front at least 10 cm from the vessel bottom and allows the response to stabilize. A small, unsuitable container can influence the test.
A transparency disk is not an NTU calibration standard for this instrument. Turbidity is also not automatically a suspended-solids concentration in mg/L. If a project requires solids mass or a site-specific turbidity-to-solids relationship, define that separate measurement and comparison rather than relabelling the turbidity field.

A radar flowmeter can combine a surface-velocity observation, surface ranging and a calculation using channel information. Confirm the measurement section and configured geometry before treating its output as the water delivered through every branch of a recirculating system. Surface velocity, water level and volumetric flow are different quantities.
A known pump command is not a flow reference. For a comparison, define the location, time interval and appropriate reference method. Account for diversions or changing storage between measurement points. A flow result cannot establish dissolved oxygen, fish condition or successful treatment by itself.

An overhead radar or ultrasonic instrument normally measures distance to the surface, which is converted using an installation reference. When water rises, that distance generally decreases. Check the sign convention, zero reference and receiving units before changing a calibration or display offset.
A level value from a supply tank, pond or treatment chamber should retain its location. One point does not establish all pond depths. Where a pump or valve uses that value, test the controller's interpretation separately from the level measurement, including how it handles an unavailable or stale reading.

A submersible pressure probe derives level from the liquid column above its sensing point. Check its position and the applicable liquid-density assumption. Where a vented cable supplies the atmospheric reference, keep the vent path dry and unobstructed; sealing the wrong part of that path can change the observation.
Likewise, an IP68 water-quality probe does not make every cable joint, controller or power connection suitable for immersion. Plan a dry service and termination location appropriate to the equipment. The calibration reference and work on removed probes should be kept separate from live control inputs during the service activity.

The NBL-WQ-EC conductivity sensor has several range configurations. Select the standard for the ordered range and temperature basis. Its slope procedure uses a standard between 10% of full scale and full scale, with the electrode vertical and at least 2 cm from the container bottom and walls. Confirm the procedure for the actual electrode version.
Conductivity, TDS and salinity fields need their own definitions. An EC check does not identify each dissolved salt or contaminant. A controller displaying conductivity also does not establish where the calibration coefficients are stored: check whether adjustment occurs in the probe, controller or receiving software so a correction is not applied twice.

The NBL-WQ-DO fluorescence sensor uses an oxygen-depleted reference for zero calibration and an appropriate air-saturation reference for slope calibration. Ordinary saturated saltwater is not a zero-oxygen reference. Follow the specified preparation and handling procedure; a casual salt addition does not establish zero dissolved oxygen.
Its procedure allows air-saturated water or the specified water-saturated-air arrangement after stabilization. Keep bubbles and deposits off the optical surface and check temperature and configured salinity compensation. The compensation setting is not an independently measured salinity channel.
| Parameter | NBL-WQ-DO | NBL-WQ-PH |
|---|---|---|
| Measurement range | 0-20.00 mg/L | 0-14 pH |
| Accuracy | +/-2% FS; temperature +/-0.3 C | +/-0.1 pH; temperature +/-0.3 C |
| Supply | 12-24 VDC | 12-24 VDC |
| Output | RS485 Modbus RTU | RS485 Modbus RTU |
| Operating temperature | 0-45 C | 0-65 C |
| Probe protection | IP68 | IP68 |
| Power and standard cable | Less than 0.3 W at 12 V; 5 m cable | 0.2 W at 12 V; 5 m cable |
The DO full-scale term corresponds to +/-0.4 mg/L over the 20 mg/L span; its 0.01 mg/L resolution is a different specification. Compare a verification difference with the applicable reference, conditions and project requirement rather than assuming the last display digit is the accuracy limit.
The NBL-WQ-CL HClO sensor belongs in a defined disinfection or treated-water sampling arrangement when that parameter is required. Its matching flow cell requires 30-60 L/h, below 0.1 MPa and within 5-50 C. A beaker or a directly immersed pond installation is not equivalent to that specified flow-cell condition.
Use a reference method that matches the HClO reporting basis. Chlorine-free water is used for the zero check; a known reference in the appropriate flow arrangement is needed for the slope procedure. Standard preparation and concentration determination require competent personnel. The instrument's calibration range is not an acceptable exposure limit for fish or shrimp.

Not automatically. Identify the model, inspect its condition and perform an appropriate verification first. Change calibration when required by the procedure or acceptance result, not merely because a communication connection has been completed.
No. It does not establish zero dissolved oxygen. Use the specified oxygen-depleted reference and the actual model's procedure, then use the appropriate saturation reference for the slope step.
Only if the instrument procedure and settings support that buffer set. The NBL-WQ-PH procedure described here uses 6.86 with the applicable 4.00 or 9.18 slope reference. Match the reference values and temperature, not just the bottle's general description.
A transparency-disk observation is not the required NTU calibration reference. Use the specified solutions and vessel geometry for the NBL-WQ-TS and keep that result separate from transparency or solids-mass measurements.
No. It verifies a measurement under defined conditions. The controller, command output and actual equipment response require separate checks, including the planned behaviour during sensor service or data loss.
Send NiuBoL the sensor model and range, reference identity, reference temperature, before-and-after-cleaning readings, installation conditions and receiving-system settings. Distinguish cleaning, comparison and calibration changes so the cause can be assessed.

Commission aquaculture equipment sensors with a method appropriate to each quantity and model. Keep pH buffers, DO references, turbidity standards, conductivity references and hydraulic comparisons distinct. Verify the measurement before changing calibration, then separately test how the monitoring and control system uses it.
Prev:Monitoring Hydrometeorological Hazards: Rainfall and River Response
Next:Pyranometer Selection: Measurement Plane, Detector and Performance Evidence
Related recommendations
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related products
Combined air temperature and relative humidity sensor
TDR Soil Moisture & Temperature Sensor for irrigation| RS485 |4-20mA| NBL-S-THR
Soil pH sensor RS485 soil Testing instrument soil ph meter for agriculture |NBL-S-PH
Wind Speed sensor Output Modbus/RS485/Analog/0-5V/4-20mA
Tipping bucket rain gauge for weather monitoring auto rainfall sensor RS485/Outdoor/stainless steel
Pyranometer Solar Radiation Sensor 4-20mA/RS485
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)