— Blogs —
—Products—
Consumer hotline +8618073152920 WhatsApp:+8615367865107
Address:Room 102, District D, Houhu Industrial Park, Yuelu District, Changsha City, Hunan Province, China
Product knowledge
Time:2026-05-26 11:48:10 Popularity:555
Quick Answer: NBL-WQ-TS applications include intake water, treatment basins and industrial water monitoring when the selected range, accuracy and installation conditions fit the measurement point. The basic probe uses infrared, 90-degree scattered-light measurement. A low displayed value is not evidence that it can verify a sub-1 NTU acceptance limit: resolution and accuracy are different specifications.

Turbidity indicates how a water sample interacts with the instrument's light, rather than directly reporting suspended-solids mass or identifying a contaminant. At an intake, a change can prompt inspection of incoming water. Across a clarification process, trends can indicate a change in solids carryover. Neither task automatically qualifies the probe for final drinking-water compliance reporting.
Choose the monitoring point before the range. Ask what action follows a change, what background value is expected and what size of change must be distinguishable. A sensor suitable for tracking a large storm-related increase may be unsuitable for deciding whether a very low finished-water limit has been met.

| Parameter | Basic NBL-WQ-TS | Engineering implication |
|---|---|---|
| Range / resolution | 0-20.00 NTU / 0.01 NTU; 0-100.0 NTU / 0.1 NTU; 0-1000.0 NTU / 0.1 NTU | Select the ordered range for normal conditions and credible excursions. |
| Turbidity accuracy | 20 NTU range: +/-3% or +/-1.5 NTU; 100 NTU range: +/-3% or +/-2 NTU; 1000 NTU range: +/-5% or +/-3 NTU | Clarify the percentage basis and application of the two terms before setting an acceptance tolerance. |
| Temperature accuracy / compensation | +/-0.3 degrees C; Pt1000 | Temperature compensation does not remove sample color, bubbles or fouling. |
| Operating conditions | 0-50 degrees C; pressure below 0.2 MPa | Check both normal operation and temporary exposure. |
| Supply / power | 12-24 V DC; 0.2 W at 12 V | Verify voltage at the probe under load, not only at the cabinet. |
| Output | RS485 Modbus RTU; 4-20 mA also listed for the product | Confirm the supplied output arrangement and wiring before ordering. |
| Housing / protection | POM and ABS; IP68 | Check chemical exposure separately; IP68 is not an unrestricted process-pressure rating. |
| Mounting / cable | Immersion, 3/4 NPT thread; standard 5 m cable | Provide support and waterproof terminations. Specify the jacket material for the site's exposure rather than assuming it. |
The basic NBL-WQ-TS product specifications differ from those of the TS-4 variants. The TS-4S shown below is a separate assembly; its range options and mechanical details must not be transferred to a basic TS order.

At an intake, select a range that covers routine water and anticipated storm events. Install where the probe sees exchanged water rather than a quiet pocket of settled solids. At a filter outlet or final-water point, first compare the required low-level accuracy with the specification above. A 0.01 NTU resolution does not cancel the absolute accuracy term or establish a regulatory method approval.
A clarification outlet may provide a useful turbidity trend, but aeration bubbles, floating material and deposits can dominate an optical reading. Keep the measurement region outside direct bubble plumes. A high-solids mixed-liquor stream is not automatically covered by the 1000 NTU range, and turbidity must not be converted to mg/L suspended solids without a validated, site-specific relationship.

Check seasonal water depth, scour, debris and access for cleaning. A position that is submerged in wet weather may become exposed during low flow; a bed-mounted probe may be buried during deposition. Record the position and immersion condition alongside the trend so a change in water level is not mistaken for a change in water quality.
Select a defined depth and a mounting arrangement that can be retrieved. Measurements at different depths can reflect different water layers, so relocation breaks direct trend comparability. Keep the optical region clear of the support structure and inspect for biological growth. A remote connection reduces travel for data collection, not the need for physical inspection.

Check temperature, pressure, color and chemical compatibility before choosing an immersion point or a conditioned sample. Cooling or depressurizing a sample may make the hardware suitable but can also change bubbles or precipitation. Where the process exceeds the probe envelope, design an appropriate sampling arrangement or select another instrument rather than relying on its IP68 label.
Keep more than 5 cm clearance from a side wall and more than 10 cm from the bottom for the basic TS installation. Provide sufficient space around the measurement region, prevent cable tension and keep the probe away from impact. Verify the position at the site's lowest operating water level.
Do not turn a nominal immersion-depth statement into permission for every pressurized installation. The operating limit remains below 0.2 MPa. The separately illustrated TS-4S assembly has its own mounting and service requirements; a picture of that model does not establish that a basic TS includes the same cleaning hardware.

On a shared RS485 Modbus RTU network, assign a unique address and match serial settings, register type, scaling and units to the supplied protocol. Compare the PLC or data logger value with a direct instrument reading. Display communication loss distinctly instead of holding a last value that appears current.
For comparison against a portable or laboratory meter, use a representative sample and record timing, optical method and handling. The guide to NTU, FNU, FTU and FAU explains why matching labels alone does not prove measurement equivalence. Check the optical window and bubbles before changing calibration; repeated calibration cannot repair an unsuitable sample location.

Do not select it from resolution alone. Compare the required tolerance, method and approval with its stated accuracy, including the absolute NTU term. A separate low-level instrument may be required even though the basic TS displays values below 1 NTU.
No. Establish the actual range and sample conditions. Dense mixed liquor, bubbles and optical saturation may require another measurement arrangement; a turbidity reading is also not a direct solids-mass result.
No. Confirm the full model, range, housing, mounting, cleaning arrangement and interface. The TS-4A image above and the multiparameter system below represent different assemblies, not additional functions included with every basic TS probe.
Inspect more frequently during commissioning and compare readings before and after permitted cleaning. Set an interval from observed fouling at that location. Do not apply one fixed interval to clean intake water, an algae-prone reservoir and a wastewater basin.
Provide the monitoring point, expected and maximum turbidity, required tolerance, water chemistry, temperature, pressure, mounting, cable route and receiving interface. These define whether a basic probe, another TS version or a different measurement approach is appropriate.

NBL-WQ-TS applications should be selected by measurement duty, not by the broad label of water-quality monitoring. Match range and accuracy to the operating decision, keep the optical region representative and check the exact model before specifying installation or maintenance. NiuBoL can review those conditions with the required output and mounting arrangement before a configuration is ordered.
NBL-WQ-CL Water Quality Sensor Online Residual Chlorine Sensor.pdf
NBL-WQ-DO Online Fluorescence Dissolved Oxygen Sensor.pdf
NBL-WQ-NHN Ammonia Nitrogen Water Quality Sensor.pdf
NBL-WQ-COD Online Water Quality COD Sensor.pdf
NBL-WQ-PH Online pH Water Quality Sensor.pdf
NBL-WQ-EC water quality conductivity sensor.pdf
NBL-WQ-BOD-4A Online BOD Sensor.pdf
NBL-WQ-TS Online Turbidity Water Quality Sensor.pdf
Prev:pH Scale, Buffer Preparation and Online Sensor Calibration
Next:Turbidity and MLSS Monitoring: Match Sensors to Treatment Stages
Related recommendations
Sensors & Weather Stations Catalog
Agriculture Sensors and Weather Stations Catalog-NiuBoL.pdf
Weather Stations Catalog-NiuBoL.pdf
Agriculture Sensors Catalog-NiuBoL.pdf
Water Quality Sensor Catalog-NiuBoL.pdf
Related products
Combined air temperature and relative humidity sensor
TDR Soil Moisture & Temperature Sensor for irrigation| RS485 |4-20mA| NBL-S-THR
Soil pH sensor RS485 soil Testing instrument soil ph meter for agriculture |NBL-S-PH
Wind Speed sensor Output Modbus/RS485/Analog/0-5V/4-20mA
0.2mm Tipping Bucket Rain Gauge for Weather & Hydrology | NiuBoL
Pyranometer Solar Radiation Sensor 4-20mA/RS485
Screenshot, WhatsApp to identify the QR code
WhatsApp number:+8615367865107
(Click on WhatsApp to copy and add friends)